Lucarelli and colleagues in this issue of Journal of Diabetes Science and Technology describe the results of interference testing for a continuous glucose monitoring system. The authors follow the Clinical Laboratory Standards Institute guideline EP7-A2, including their conclusions, in which the concepts of a statistically significant interfering substance and a clinically important interference have been combined in a way whereby information from the experiment has been lost and could be misleading. A better way to treat the data is presented, including a simulation method to evaluate the effects of interferences.
机译:Lucarelli及其同事在本期《糖尿病科学与技术杂志》上描述了连续葡萄糖监测系统的干扰测试结果。作者遵循临床实验室标准协会的指南EP7-A2,包括其结论,其中统计学上显着的干扰物质和临床上重要的干扰的概念已结合在一起,从而丢失了来自实验的信息并可能产生误导。提出了一种更好的数据处理方法,包括一种评估干扰影响的仿真方法。