Background: MARD (mean absolute relative difference) is increasingly used to describe performance of glucose monitoring systems, providing a single-value quantitative measure of accuracy and allowing comparisons between different monitoring systems. This study reports MARDs for the OneTouch Verio? glucose meter clinical data set of 80 258 data points (671 individual batches) gathered as part of a 7.5-year self-surveillance program Methods: Test strips were routinely sampled from randomly selected manufacturer’s production batches and sent to one of 3 clinic sites for clinical accuracy assessment using fresh capillary blood from patients with diabetes, using both the meter system and standard laboratory reference instrument. Results: Evaluation of the distribution of strip batch MARD yielded a mean value of 5.05% (range: 3.68-6.43% at ±1.96 standard deviations from mean). The overall MARD for all clinic data points (N = 80 258) was also 5.05%, while a mean bias of 1.28 was recorded. MARD by glucose level was found to be consistent, yielding a maximum value of 4.81% at higher glucose (≥100 mg/dL) and a mean absolute difference (MAD) of 5.60 mg/dL at low glucose (<100 mg/dL). MARD by year of manufacture varied from 4.67-5.42% indicating consistent accuracy performance over the surveillance period. Conclusions: This 7.5-year surveillance program showed that this meter system exhibits consistently low MARD by batch, glucose level and year, indicating close agreement with established reference methods whilste exhibiting lower MARD values than continuous glucose monitoring (CGM) systems and providing users with confidence in the performance when transitioning to each new strip batch.
机译:背景:MARD(绝对绝对相对差)越来越多地用于描述葡萄糖监测系统的性能,提供单值定量的准确性度量,并允许在不同监测系统之间进行比较。这项研究报告了OneTouch Verio的MARD?血糖仪临床数据集(共7.5年自我监测计划的一部分)收集了80258个数据点(671个独立批次)方法:定期从随机选择的制造商生产批次中抽取试纸,并送至3个临床站点之一进行临床使用仪表系统和标准实验室参考仪器,使用来自糖尿病患者的新鲜毛细血管血进行准确性评估。结果:对带状批料MARD的分布进行评估得出平均值为5.05%(范围为3.68-6.43%,平均值的平均值为±1.96)。所有临床数据点(N = 80 258)的总体MARD也是5.05%,而平均偏差为1.28。发现葡萄糖水平的MARD是一致的,在较高葡萄糖(≥100mg / dL)时,最大值为4.81%,在较低葡萄糖(<100 mg / dL)时,平均绝对差(MAD)为5.60 mg / dL。 。制造年份的MARD在4.67%至5.42%之间变化,表明在整个监视期内,其准确性始终如一。结论:这项为期7.5年的监视程序表明,该仪表系统在批次,葡萄糖水平和年份方面始终显示出较低的MARD,表明与既定的参考方法有着高度的一致性,同时其MARD值也低于连续血糖监测(CGM)系统,并为用户带来了信心过渡到每个新带钢批次时的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号