Dosimetric validation of new semiconductor diode dosimetry system for intensity modulated radiotherapy

机译:用于强度调制放疗的新型半导体二极管剂量测定系统的剂量测定验证

获取原文
获取外文期刊封面目录资料

摘要

Introduction: The new diode Isorad was validated for intensity modulated radiotherapy (IMRT) and the observations during the validation are reported. Materials and Methods: The validation includes intrinsic precision, post-irradiation stability, dose linearity, dose-rate effect, angular response, source to surface (SSD) dependence, field size dependence, and dose calibration. Results: The intrinsic precision of the diode was more than 1% (1 σ). The linearity found in the whole range of dose analyzed was 1.93% (R 2 = 1). The minimum and maximum variation in the measured and calculated dose were found to be 0.78% (with 25 MU at ioscentre) and 4.8% (with 1000 MU at isocentre), respectively. The maximal variation in angular response with respect to arbitrary angle 0° found was 1.31%. The diode exhibited a 51.7% and 35% decrease in the response in the 35 cm and 20 cm SSD range, respectively. The minimum and the maximum variation in the measured dose from the diode and calculated dose were 0.82% (5 cm × 5 cm) and 3.75% (30 cm × 30 cm), respectively. At couch 270°, the response of the diode was found to vary maximum by 1.4% with ΁ 60 gantry angle. Mean variation between measured dose with diode and planned dose by TPS was found to be 1.3% (SD 0.75) for IMRT patient-specific quality assurance. Conclusion: For the evaluation of IMRT, use of cylindrical diode is strongly recommended.
机译:简介:新型二极管Isorad经过了强度调制放射治疗(IMRT)的验证,并报告了验证期间的观察结果。材料和方法:验证包括固有精度,辐照后稳定性,剂量线性,剂量率效应,角度响应,源对表面(SSD)依赖性,场大小依赖性以及剂量校准。结果:二极管的固有精度超过1%(1σ)。在所分析的整个剂量范围内发现的线性为1.93%(R 2 = 1)。发现测量和计算剂量的最小和最大变化分别为0.78%(ioscentre为25 MU)和4.8%(isocentre为1000 MU)。相对于任意角度0°的角度响应的最大变化为1.31%。二极管在35 cm和20 cm SSD范围内的响应分别降低了51.7%和35%。二极管测得的剂量和计算出的剂量的最小和最大变化分别为0.82%(5 cm×5 cm)和3.75%(30 cm×30 cm)。在270°的工作台上,发现二极管的响应在΁60龙门角时最大变化1.4%。对于IMRT患者特定的质量保证,发现二极管测量剂量与TPS计划剂量之间的平均差异为1.3%(SD 0.75)。结论:为了评估IMRT,强烈建议使用圆柱形二极管。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号