Cardiac-resynchronization therapy (CRT) improves the cardiac function of patients with left ventricular (LV) dyssynchrony. Maintenance of the biventricular pacing rate is very important in managing the hemodynamics in patients implanted with CRT devices. A low biventricular pacing rate, for example, in cases with atrial fibrillation or rapid intrinsic atrioventricular (AV) conduction, decreases the benefits of CRT. The LUMAX HF-T 540 device series (BIOTRONIK, Berlin, Germany) has a LV-triggered pace algorithm, which allows biventicular pacing rates to be maintained even during rapid intrinsic rhythms caused by shortened AV conduction and/or premature ventricular contraction (PVC) occurring in the right ventricle. We encountered a case of CRT device implantation with a defibrillator wherein this triggered pace algorithm caused a spike on T-wave due to T-wave oversensing. By remote monitoring, we were also able to determine that the T-wave oversensing was due to a PVC. The LUMAX 540 series allows for changes in the sensing threshold and filter settings of the device, which facilitated the elimination of T-wave oversensing in this case.
机译:心脏再同步治疗(CRT)可改善左心室(LV)不同步的患者的心脏功能。在控制植入了CRT装置的患者的血液动力学方面,维持双室起搏速度非常重要。低的双心室起搏速度,例如在房颤或快速的固有房室传导中,会降低CRT的益处。 LUMAX HF-T 540设备系列(BIOTRONIK,柏林,德国)具有LV触发的起搏算法,即使在因缩短的AV传导和/或过早的心室收缩(PVC)引起的快速固有节律期间,也可以保持双心室起搏率发生在右心室。我们遇到了一个带有除颤器的CRT设备植入案例,其中该触发起搏算法由于T波感应过度而导致T波尖峰。通过远程监控,我们还可以确定T波超感是由于PVC引起的。 LUMAX 540系列允许更改设备的感应阈值和滤波器设置,在这种情况下,这有助于消除T波过感应。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号