首页> 外文期刊>Journal of arrhythmia. >Paradoxical undersensing of atrial electrograms during atrial fibrillation due to repeated activation of the quiet timer blanking interval: Case report and review of the literature
【24h】

Paradoxical undersensing of atrial electrograms during atrial fibrillation due to repeated activation of the quiet timer blanking interval: Case report and review of the literature

机译:由于反复激活安静定时器消隐间隔而导致的心房颤动期间心电图的反常感知:病例报告和文献复习

获取原文

摘要

This report describes a case of paradoxical atrial undersensing by a dual-chamber pacemaker during paroxysmal atrial fibrillation. Undersensing of 5.6mV atrial signals at a programmed sensitivity of 0.5mV returned to normal sensing by decreasing atrial sensitivity to 1.0mV. This uncommon phenomenon can be explained by a repeated activation of the quiet timer blanking interval. Knowledge of this phenomenon is important in the current pacemaker management to improve the accuracy of the diagnostic feature for atrial tachyarrhythmia burden and to avoid unnecessary lead revisions.
机译:该报告描述了在阵发性心房颤动期间双腔起搏器发生自相矛盾的心房不足的情况。编程灵敏度为0.5mV时,对5.6mV心房信号的欠感可通过将心房灵敏度降至1.0mV而恢复到正常感测。这种不常见的现象可以通过重复激活静默计时器消隐间隔来解释。在当前的起搏器管理中,了解这一现象对于提高房性快速性心律失常负担的诊断功能的准确性并避免不必要的导线修订非常重要。