Feasibility of a remote, automated daily delivery verification of volumetric‐modulated arc therapy treatments using a commercial record and verify system

机译:使用商业记录和验证系统进行容积调制电弧疗法治疗的远程自动每日交付验证的可行性

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Volumetric‐modulated arc therapy (VMAT) is an effective but complex technique for delivering radiation therapy. VMAT relies on precise combinations of dose rate, gantry speed, and multileaf collimator (MLC) shapes to deliver intensity‐modulated patterns. Such complexity warrants the development of correspondingly robust performance verification systems. In this work, we report on a remote, automated software system for daily delivery verification of VMAT treatments. The performance verification software system consists of three main components: (1) a query module for retrieving daily MLC, gantry, and jaw positions reported by the linear accelerator control system to the record and verify system; (2) an analysis module which reads the daily delivery report generated from the database query module, compares the reported treatment positions against the planned positions, and compiles delivery position error reports; and (3) a graphical reporting module which displays reports initiated by a user anywhere within the institutional network or which can be configured to alert authorized users when predefined tolerance values are exceeded. The utility of the system was investigated through analysis of patient data collected at our clinic. Nearly 2500 VMAT fractions have been analyzed with the delivery verification system at our institution. The average percentage of reported MLC leaf positions within 3 mm, gantry positions within 2°, and jaw positions within 3 mm of their planned positions was , and , respectively. The level of agreement between planned and reported MLC positions decreased for treatment plans requiring larger MLC leaf movements between control points. Differences in the reported MLC position error between the delivery verification system and data extracted manually from the control system were noted; however, the differences are likely systematic and, therefore, may be characterized if appropriately accounted for. Further investigation is needed to confirm the utility and accuracy of the system. PACS numbers: 87.55.N‐, 87.55.T‐, 87.55.Qr
机译:容积调制电弧疗法(VMAT)是一种有效但复杂的放射治疗技术。 VMAT依靠剂量率,门架速度和多叶准直器(MLC)形状的精确组合来提供强度调制的图案。这种复杂性保证了相应强大的性能验证系统的开发。在这项工作中,我们报告了一个远程自动化软件系统,用于VMAT治疗的每日交付验证。性能验证软件系统由三个主要组件组成:(1)查询模块,用于检索线性加速器控制系统报告给记录和验证系统的每日MLC,机架和钳口位置; (2)分析模块,其读取数据库查询模块生成的每日交付报告,将所报告的治疗位置与计划位置进行比较,并编制交付位置错误报告; (3)图形报告模块,其显示由用户在机构网络内的任何地方发起的报告,或者可以被配置为在超过预定的公差值时向授权用户发出警报。通过分析我们诊所收集的患者数据,调查了该系统的实用性。我们机构的交付验证系统已对近2500个VMAT分数进行了分析。所报告的MLC叶片位置在距其计划位置3 mm以内,龙门位置在2°以内以及颚位置在3 mm以内的平均百分比分别为和。对于需要在控制点之间进行较大MLC叶片移动的处理计划,计划的和报告的MLC位置之间的一致性降低。记录了交付验证系统与从控制系统手动提取的数据之间报告的MLC位置误差的差异;但是,差异可能是系统性的,因此,如果适当考虑,就可以表征差异。需要进一步调查以确认系统的实用性和准确性。 PACS编号:87.55.N‐,87.55.T‐,87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号