Performance evaluation of the RITG148+ set of TomoTherapy quality assurance tools using RTQA2 radiochromic film

机译:使用RTQA2放射变色胶片对RITG148 + TomoTherapy质量保证工具集进行性能评估

获取原文
           

摘要

Version 6.3 of the RITG 148 + software package offers eight automated analysis routines for quality assurance of the TomoTherapy platform. A performance evaluation of each routine was performed in order to compare RITG 148 + results with traditionally accepted analysis techniques and verify that simulated changes in machine parameters are correctly identified by the software. Reference films were exposed according to AAPM TG-148 methodology for each routine and the RITG 148 + results were compared with either alternative software analysis techniques or manual analysis techniques in order to assess baseline agreement. Changes in machine performance were simulated through translational and rotational adjustments to subsequently irradiated films, and these films were analyzed to verify that the applied changes were accurately detected by each of the RITG 148 + routines. For the Hounsfield unit routine, an assessment of the “Frame Averaging” functionality and the effects of phantom roll on the routine results are presented. All RITG 148 + routines reported acceptable baseline results consistent with alternative analysis techniques, with 9 of the 11 baseline test results showing agreement of 0.1 mm/0.1° or better. Simulated changes were correctly identified by the RITG 148 + routines within approximately 0.2 mm / 0.2 ° with the exception of the Field Center vs. Jaw Setting routine, which was found to have limited accuracy in cases where field centers were not aligned for all jaw settings due to inaccurate autorotation of the film during analysis. The performance of the RITG 148 + software package was found to be acceptable for introduction into our clinical environment as an automated alternative to traditional analysis techniques for routine TomoTherapy quality assurance testing.PACS number(s): 87.55.Qr
机译:RITG 148 +软件包的6.3版提供了八个自动分析例程,以确保TomoTherapy平台的质量。为了将RITG 148 +结果与传统上接受的分析技术进行比较,并验证软件是否正确识别了机器参数的模拟变化,对每个例程进行了性能评估。根据AAPM TG-148方法对每种常规方法曝光参考胶片,并将RITG 148 +结果与替代软件分析技术或手动分析技术进行比较,以评估基线一致性。通过平移和旋转调整对随后辐照过的胶片进行模拟,以模拟机器性能的变化,并对这些胶片进行分析,以验证每个RITG 148 +例程都能准确检测出所施加的变化。对于Hounsfield单位例行程序,将评估“帧平均”功能以及幻影滚动对例行程序结果的影响。所有RITG 148 +例行程序均报告了可接受的基线结果,与替代分析技术一致,在11个基线测试结果中有9个显示出0.1 mm / 0.1°或更好的一致性。 RITG 148 +程序在大约0.2 mm / 0.2°范围内正确识别了模拟变化,但“场中心”与“下颌设置”例行程序除外,如果在所有下颌设置中未将场中心对齐,则发现精度有限由于分析过程中薄膜的自动旋转不准确。发现RITG 148 +软件包的性能可以接受,可以作为常规TomoTherapy质量保证测试的传统分析技术的自动替代方案引入我们的临床环境.PACS编号:87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号