Correction of motion‐induced misalignment in co‐registered PET/CT and MRI (T1/T2/FLAIR) head images for stereotactic radiosurgery

机译:矫正立体定向放射外科手术在PET / CT和MRI(T1 / T2 / FLAIR)头部共同成像中运动引起的未对准

获取原文

摘要

The purpose was to evaluate and correct the co‐registration of diagnostic PET/CT and MRI/MRI images for stereotactic radiosurgery (SRS) using 3D volumetric image registration (3DVIR). The 3DVIR utilizes the homogeneity of color distribution over a volumetric anatomical landmark as the registration criterion with submillimeter accuracy. Fifty‐three PET/CT and MRI (T1, T2 and FLAIR) image sets of patients with brain lesions were acquired sequentially from a hybrid PET/CT or an MRI scanner with common diagnostic head holding devices. Twenty‐five sets of head images were scanned over a 10‐minute interval and 14 whole‐body sets were scanned over a 30‐minute interval. Fourteen sets of MRI images were acquired, and each 3‐modal image set (T1, T2 and FLAIR) was scanned in sequence at time 0, and minutes. The misalignments in these “co‐registered” images were evaluated and corrected using the 3DVIR. Using the head immobilization devices commonly found in diagnostic PET/CT and MRI/MRI imaging, 80%–100% of these “co‐registered” images were identified as misaligned. For PET/CT, the magnitude of misalignment was and for 10‐minute scans, and and for 30‐minute scans. For MRI/MRI, the magnitude was and for 5‐minute scan intervals, and and for 20‐minute intervals. Small, but significant, misalignment is present in the co‐registered diagnostic PET/CT and MRI/MRI images and can be corrected in SRS treatment planning using the volumetric image registration for improved target localization within the clinical error tolerance. PACS numbers: 87.53.Ly, 87.57.nj, 87.57.uk, 87.57.Q‐, 87.61.jc, 87.19.xc Conflict of Interest Statement: The authors do not have any conflict of interest on this research report.
机译:目的是评估和纠正使用3D立体图像配准(3DVIR)的立体定向放射外科(SRS)的诊断PET / CT和MRI / MRI图像的共配准。 3DVIR利用体积解剖界标上颜色分布的均匀性作为亚毫米级精度的配准标准。从具有常规诊断头固定装置的混合PET / CT或MRI扫描仪中顺序采集脑损伤患者的53张PET / CT和MRI(T1,T2和FLAIR)图像集。在10分钟的间隔内扫描了25组头部图像,在30分钟的间隔内扫描了14个全身图像。采集了14组MRI图像,并分别在时间0和分钟数扫描了每个3模态图像集(T1,T2和FLAIR)。使用3DVIR对这些“共同注册”图像中的未对准进行了评估和纠正。使用通常在诊断性PET / CT和MRI / MRI成像中使用的头部固定装置,可以将这些“共同注册”图像中的80%–100%识别为未对准。对于PET / CT,未对准的大小分别为10分钟和30分钟。对于MRI / MRI,幅值分别为,5分钟扫描间隔和20分钟间隔。共同记录的诊断性PET / CT和MRI / MRI图像中存在微小但明显的失准,并且可以在SRS治疗计划中使用体积图像注册进行校正,以在临床误差容忍度内改善目标定位。 PACS编号:87.53.Ly,87.57.nj,87.57.uk,87.57.Q-,87.61.jc,87.19.xc利益冲突声明:作者对此研究报告没有任何利益冲突。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号