首页> 外文期刊>Journal of applied clinical medical physics / >Verification measurements of an eMC algorithm using a 2D ion chamber array
【24h】

Verification measurements of an eMC algorithm using a 2D ion chamber array

机译:使用2D离子室阵列的eMC算法的验证测量

获取原文
           

摘要

The aim of this study was to assess the suitability of the Im' RT MatriXX 2D ion chamber array for performing verification measurements on the Varian Eclipse electron Monte Carlo (eMC) algorithm for a range of clinical energies (6, 12, and 20 MeV) on a Varian 2100iX linear accelerator. Firstly, the suitability of the MatriXX for measuring percentage depth doses (PDD) in water was assessed, including characterization of the inherent buildup found in the MatriXX. Secondly the suitability of the MatriXX for measuring dose distributions in homogeneous and heterogeneous phantoms was assessed using gamma analysis at 3 % / 3 mm . It was found that after adjusting the PDD curves for the inherent buildup, that the position of R 50 , D measured using the MatriXX agreed to within 1 mm to the PDDs generated using the eMC algorithm for all energies used in this study. Gamma analysis at 3 % / 3 mm showed very good agreement ( > 95 % ) for all cases in both homogeneous and heterogeneous phantoms. It was concluded that the Im' RT MatriXX is a suitable device for performing eMC verification and could potentially be used for routine energy checks of electron beams.PACS number(s): 87.55.km, 87.55.Qr
机译:这项研究的目的是评估Im'RT MatriXX 2D离子室阵列在Varian Eclipse电子蒙特卡罗(eMC)算法上对一系列临床能量(6、12和20 MeV)进行验证测量的适用性。在Varian 2100iX线性加速器上。首先,评估了MatriXX在水中测量百分比深度剂量(PDD)的适用性,包括表征了在MatriXX中发现的固有堆积。其次,使用3%/ 3 mm的伽马分析评估了MatriXX在均质和异质体模中测量剂量分布的适用性。结果发现,在针对固有积累调整PDD曲线后,对于本研究中使用的所有能量,使用MatriXX测量的R 50,D位置与使用eMC算法生成的PDD的位置一致,在1 mm之内。在均质体模和异质体模中,所有情况下的3%/ 3mm伽马分析都显示出非常好的一致性(> 95%)。结论是Im'RT MatriXX是执行eMC验证的合适设备,可潜在地用于电子束的常规能量检查.PACS编号:87.55.km,87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号