Evaluation of the Technicon Axon analyser

机译:Technicon Axon分析仪的评估

获取原文
       

摘要

An evaluation of the Technicon Axon analyser was carried out following the guidelines of the ‘Sociedad Espa?ola de Química Clínica’ and the European Committee for Clinical Laboratory Standards.A photometric study revealed acceptable results at both 340 nm and 404 nm. Inaccuracy and imprecision were lower at 404 nm than at 340 nm, although poor dispersion was found at both wavelengths, even at low absorbances. Drift was negligible, the imprecision of the sample pipette delivery system was greater for small sample volumes, the reagent pipette delivery system imprecision was acceptable and the sample diluting system study showed good precision and accuracy.Twelve analytes were studied for evaluation of the analyser under routine working conditions. Satisfactory results were obtained for within-run imprecision, while coefficients of variation for betweenrun imprecision were much greater than expected. Neither specimenrelated nor specimen-independent contamination was found in the carry-over study. For all analytes assayed, when comparing patient sample results with those obtained in a Hitachi 737 analyser, acceptable relative inaccuracy was observed.
机译:根据“西班牙社会科学杂志”和欧洲临床实验室标准委员会的指导方针,对Technicon Axon分析仪进行了评估。光度学研究表明,在340 nm和404 nm处均可接受的结果。尽管在404 nm和340 nm处均存在较差的色散,即使在低吸光度下,在两个波长下均发现较差的色散,但其准确性和不精确度均低于340 nm。漂移可以忽略不计,小体积样品移液管的不精确度更大,试剂移液管的不精确度可以接受,样品稀释系统的研究显示出良好的准确性和准确性。在常规条件下对十二种分析物进行分析仪评估工作环境。运行内不精确度获得了令人满意的结果,而运行间不精确度的变异系数则远大于预期。在残留研究中未发现与标本有关或与标本无关的污染。对于所有测定的分析物,将患者样品结果与在Hitachi 737分析仪中获得的结果进行比较时,观察到可接受的相对误差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号