首页> 外文期刊>Journal of applied clinical medical physics / >Angular dependence of the output of a kilovoltage X‐ray therapy unit
【24h】

Angular dependence of the output of a kilovoltage X‐ray therapy unit

机译:千伏X射线治疗仪输出的角度依赖性

获取原文
获取外文期刊封面目录资料

摘要

During the recommissioning of a Philips RT‐250 kilovoltage X‐ray unit, unexpected output variations with tube head rotation (cross‐plane) and tube head tilt (in‐plane) were observed. The measured output showed an increase of up to 7.3% relative to the neutral position (0° in‐plane and 0° cross‐plane) over the possible range of angles of in‐plane rotation for 75 kVp (half‐value layer, ). A less pronounced but noticeable output change (with respect to the neutral position) was observed for cross‐plane rotation reaching 2% for the 225 kVp beam ( ). This output variation was observed while manually adjusting the current to maintain constancy according to the current meter gauge. In order to address the observed output dependence with head orientation, the dose rate monitor chamber of the kilovoltage unit was calibrated to monitor the beam output in real time. The dose rate was manually adjusted to maintain a constant dose rate (in r/min) as displayed on the r/min gauge. This approach resulted in maintaining beam output for the 75 kVp and the 225 kVp beams within for the in‐plane angle variation and for the cross‐plane angle variation. A daily output check that includes ion chamber‐based measurements at the neutral position and an in‐plane angle of 45° has been implemented using the constant dose rate approach to monitor the stability of the X‐ray beams. As a result of the output variations with in/cross‐plane rotation, the quality control (QC) procedures that are typically used for clinical setup have been modified to test the stability of the beams under the non‐neutral positioning of the X‐ray tube. PACS number: 87.56.Fc
机译:在重新调试Philips RT-250千伏X射线装置的过程中,观察到输出随管头旋转(跨平面)和管头倾斜(面内)而产生的意外变化。在75 kVp的平面内旋转角度范围内(半值层,),所测得的输出相对于中性位置(平面内0°和横向平面0°)增加了7.3%。 。对于225 kVp光束(),跨面旋转达到2%时,观察到的输出变化(相对于中立位置)的变化较小,但很明显。当根据电流表规格手动调节电流以保持恒定时,观察到此输出变化。为了解决观察到的与头部方向有关的输出依赖性,千伏单位的剂量率监测室经过校准以实时监测电子束输出。手动调整剂量率,以保持恒定的剂量率(以r / min为单位),如r / min计上所示。这种方法导致对于面内角度变化和跨面角度变化,将75 kVp和225 kVp光束的输出保持在该范围内。每天的输出检查包括在中性位置基于离子室的测量以及45°的平面内角度,这是使用恒定剂量率方法进行的,以监视X射线束的稳定性。由于输出在平面内/跨平面旋转中的变化,因此修改了通常用于临床设置的质量控制(QC)程序,以在X射线非中性定位下测试光束的稳定性管。 PACS编号:87.56.Fc

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号