A quality assurance procedure to evaluate cone‐beam CT image center congruence with the radiation isocenter of a linear accelerator

机译:评估锥形束CT图像中心与线性加速器的辐射等角点的一致性的质量保证程序

获取原文
       

摘要

A quality assurance (QA) procedure was developed to evaluate the congruence between the cone‐beam computed tomography (CBCT) image center and the radiation isocenter on a Varian Trilogy linac. In contrast to the published QA procedures, this method did not require a ball bearing (BB) phantom to be placed exactly at the radiation isocenter through precalibrated room lasers or light field crosshairs. The only requirement was that the BB phantom be in a stationary position near the radiation isocenter during the image acquisition process. The radiation isocenter was determined with respect to the center of the BB using a Winston‐Lutz test. The CBCT image center was found to have excellent short‐term positional reproducibility (i.e., less than 0.1 mm of wobble in each of the x (lateral), y (vertical), and z (longitudinal) directions) in 10 consecutive acquisitions. Measured over a seven‐month period, the CBCT image center deviated from the radiation isocenter by , and . The z displacement of the 3D CBCT image center was highly correlated with that of the 2D kV portal image center. The correlation coefficients in the x and y directions were poor ( and ‐0.35, respectively). Systematic discrepancies were found between the CBCT image center and the 2D MV, kV portal image centers. For the linear accelerator studied, we detected a 0.8 mm discrepancy between the CBCT image center and the MV EPID image center in the anterior‐posterior direction. This discrepancy was demonstrated in a clinical case study where the patient was positioned with CBCT followed by MV portal verification. The results from the new QA procedure are useful for guiding high‐precision patient positioning in stereotactic body radiation therapy. PACS number: 87.55.Qr
机译:制定了质量保证(QA)程序,以评估Varian Trilogy直线加速器上的锥束计算机断层扫描(CBCT)图像中心和辐射等中心线之间的一致性。与已发布的QA程序相反,此方法不需要通过预先校准的室内激光器或光场十字准线将球体(BB)体模精确地放置在辐射等角点处。唯一的要求是,在图像采集过程中,BB体模必须位于辐射等角点附近的固定位置。辐射等中心线是通过Winston-Lutz测试相对于BB中心确定的。发现CBCT图像中心在10次连续采集中具有出色的短期位置可重复性(即,x(横向),y(垂直)和z(纵向)方向的摆动均小于0.1毫米)。经过七个月的测量,CBCT影像中心偏离了辐射等中心线,且和。 3D CBCT图像中心的z位移与2D kV门户图像中心的z位移高度相关。 x和y方向的相关系数很低(分别为-0.35)。发现CBCT影像中心与2D MV,kV门户影像中心之间存在系统差异。对于所研究的线性加速器,我们在前后方向上检测到CBCT图像中心与MV EPID图像中心之间存在0.8 mm的差异。这种差异在临床案例研究中得到了证明,在该案例中,对患者进行了CBCT定位,然后进行了MV门户验证。新的QA程序的结果可用于指导立体定向人体放射治疗中的高精度患者定位。 PACS编号:87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号