Validation of Platelet Counting Accuracy With the Celltac F AutomatedHematology Analyzer

机译:Celltac F自动血液分析仪对血小板计数准确性的验证

获取原文

摘要

Rapid and accurate analysis of platelet count plays an important role in evaluating hemorrhagic status. Therefore, we evaluated platelet counting performance of a hematology analyzer, Celltac F (MEK-8222, Nihon Kohden Corporation, Tokyo, Japan), thatfeatures easy use with low reagent consumption and high throughput while occupying minimal space in the clinical laboratory. All blood samples were anticoagulated with dipotassium ethylenediaminetetraacetic acid (EDTA-2K). The samples were storedat room temperature (18°C–22°C) and tested within 4 hours of phlebotomy. We evaluated the counting ability of the Celltac F hematology analyzer by comparing it with the platelet counts obtained by the flow cytometry method that ISLH and ICSH recommended, and also the manual visual method by Unopette (Becton Dickinson Vacutainer Systems). The ICSH/ISLH reference method is based on the fact that platelets can be stained with monoclonal antibodies to CD41 and/or CD61. The dilution ratio was optimized after the precision, coincidence events, and debris counts were confirmed by the reference method. Good correlation of platelet count between the Celltac F and the ICSH/ISLH reference method (r=0.99), and the manual visual method (r=0.93) were obtained. The regressions werey=0.90x+9.0andy=1.11x+8.4, respectively. We conclude that the Celltac F hematology analyzer for platelet counting was well suited to the ICSH/ISLH reference method for rapidness and reliability.
机译:快速准确的血小板计数分析在评估出血状态中起着重要作用。因此,我们评估了血液分析仪Celltac F(MEK-8222,日本东京日本Nihon Kohden公司)的血小板计数性能,该产品具有易于使用,试剂消耗低和通量高的特点,同时在临床实验室中占用的空间很小。所有血样均用乙二胺四乙酸二钾(EDTA-2K)抗凝。样品在室温(18°C–22°C)下保存,并在放血4小时内进行测试。我们通过将Celltac F血液分析仪的计数能力与ISLH和ICSH推荐的流式细胞仪方法以及Unopette(Becton Dickinson Vacutainer Systems)的手动目测方法进行比较,评估了其计数能力。 ICSH / ISLH参考方法基于以下事实:血小板可用抗CD41和/或CD61的单克隆抗体染色。通过参考方法确定精密度,同时发生的事件和碎片计数后,可以优化稀释比。 Celltac F与ICSH / ISLH参考方法(r = 0.99)和手动视觉方法(r = 0.93)之间的血小板计数具有良好的相关性。回归分别为y = 0.90x + 9.0和y = 1.11x + 8.4。我们得出的结论是,用于血小板计数的Celltac F血液分析仪非常适合于ICSH / ISLH参考方法的快速性和可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号