Introduction: Body fluid (BF) analysis is critical to the diagnosis and monitoring of several pathological conditions. The limitations of manual cell counts have led to greater interest in the development of automated BF analysis. Objective: To evaluate the analytical performance of the Sysmex XE-5000 hematology analyzer in the analysis of pleural and ascitic fluids in the laboratory routine of a large university hospital. Methods: A total of 56 samples (35 ascitic and 21 pleural fluids) were analyzed by manual optical microscopy (OM) and XE-5000. Analytical performance includes linearity, carryover, functional sensitivity and comparison of patient samples. Results: Performance studies showed linearity up to 25,825 WBC-BF/?μl ( r 2 = 0.999), WBC-BF showed carryover of 0.18%, and the lower limit of quantitation was set at 22 WBC-BF/?μl. Good correlations between the methods were observed just for total cell (TC-BF) and white blood cell (WBC-BF) counts in pleural and ascitic fluids. The high-fluorescence cell count (HF-BF) showed poor correlation but high positive predictive value (PPV) for both fluids (94.74% for pleural and 96.97% for ascitic fluid). Conclusion: XE-5000 provides accurate and precise count for TC-BF, WBC-BF and polymorphonuclear cells (PMN-BF) in pleural and ascitic fluids in medical decision levels, but the morphological differentiation should continue to be held by OM. Histogram and scattergram displayed on XE-5000 must always be analyzed to assess if there is any interference or flag. The HF-BF parameter is a potential tool for screening.
机译:简介:体液(BF)分析对于多种病理状况的诊断和监测至关重要。手动细胞计数的局限性引起了对自动BF分析开发的更大兴趣。目的:评估Sysmex XE-5000血液分析仪在大型大学医院实验室常规分析胸水和腹水中的分析性能。方法:采用手动光学显微镜(OM)和XE-5000对56份样本(35份腹水和21份胸水)进行了分析。分析性能包括线性,残留,功能灵敏度和患者样品比较。结果:性能研究显示线性高达25,825 WBC-BF /?μl(r 2 = 0.999),WBC-BF残留度为0.18%,定量下限设定为22 WBC-BF /?μl。仅对于胸膜和腹水中的总细胞(TC-BF)和白细胞(WBC-BF)计数,方法之间观察到良好的相关性。两种液体的高荧光细胞计数(HF-BF)均显示出较低的相关性,但阳性预测值(PPV)较高(胸膜为94.74%,腹水为96.97%)。结论:XE-5000可在医学决策水平上对胸膜和腹水中的TC-BF,WBC-BF和多形核细胞(PMN-BF)进行准确而精确的计数,但OM应继续保持形态分化。必须始终分析XE-5000上显示的直方图和散点图,以评估是否存在干扰或标记。 HF-BF参数是筛选的潜在工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号