首页> 外文期刊>Diving and hyperbaric medicine: the journal of the South Pacific Underwater Medicine Society >Performance of the BBraun Perfusor Space syringe driver under hyperbaric conditions
【24h】

Performance of the BBraun Perfusor Space syringe driver under hyperbaric conditions

机译:高压环境下BBraun Perfusor Space注射器驱动器的性能

获取原文

摘要

Background The BBraun Perfusor Space? syringe driver is already in use by ambulance services and retrieval teams but has not previously been assessed for hyperbaric chamber use. Methods Pump flow accuracy was tested at rates between 1 and 40 ml·hr?1 using three different brands of 50 ml syringe. Function of the occlusion alarms was assessed using the same syringes. The hyperbaric profile involved pressurisation to 284 kPa at 30 kPa·min?1, 30 min at 284 kPa and decompression at 30 kPa·min?1. Output was recorded from differences in weight of collection containers. A single device was tested. Results Performance was highly dependent on the syringe type used, with two of the three 50 ml syringes used demonstrating 'stiction' at both low and high occlusion pressure alarm settings, most marked during pressurisation. On decompression from 284 kPa all syringes alarmed at significantly lower pressures. Because of the stiction problems only the flow measurements for the BBrown Omnifix 50 ml syringes are reported. At a pressure of 284 kPa, the difference between programmed and delivered rates was within the manufacturer's specification of 10%: at 40 ml·h?1 (median variation 1.25%, IQR 0.5?1.7%), 10 ml·h?1 (8.6%, IQR 8?9.2%), 5 ml·h?1(-8.8%, IQR -1.6?8.8%) and 1 ml·h?1 (-4%, IQR 4?12%). Pressurisation was associated with significantly lower flow rates whilst decompression was associated with significantly increased rates. Limited testing at 405 kPa was also within the manufacturer’s specifications. Conclusion A BBraun Infusor Space syringe driver performed within acceptable performance criteria but is highly dependent on syringe type and flow rates. The potential for the device to under deliver on pressurisation and over deliver on depressurisation, however, suggests vigilance and appropriate rate adjustments may be necessary during these phases.
机译:背景BBraun香水空间?救护车服务和取回小组已经在使用注射器驱动程序,但以前尚未对高压舱的使用进行过评估。方法使用三个不同品牌的50 ml注射器以1至40 ml·hr?1的速率测试泵的流量精度。使用相同的注射器评估阻塞警报的功能。高压曲线包括在30 kPa·min?1下加压至284 kPa,在284 kPa时加压30分钟和在30 kPa·min?1下减压。从收集容器的重量差异记录输出。测试了单个设备。结果性能在很大程度上取决于所使用的注射器类型,在低和高咬合压力警报设置下使用的三个50 ml注射器中,有两个显示出“粘着”,在加压过程中最明显。从284 kPa减压时,所有注射器均会在明显更低的压力下发出警报。由于粘滞问题,仅报告了BBrown Omnifix 50 ml注射器的流量测量结果。在284 kPa的压力下,编程速率和输送速率之间的差异在制造商规定的10%范围内:在40 ml·h?1(中位数变化为1.25%,IQR为0.5?1.7%),10 ml·h?1( 8.6%,IQR 8%〜9.2%),5 ml·h?1(-8.8%,IQR -1.6%8.8%)和1 ml·h?1(-4%,IQR 4%〜12%)。加压与流速显着降低相关,而减压与流速显着相关。 405 kPa的受限测试也符合制造商的规格。结论BBraun Infusor Space注射器驱动器在可接受的性能标准内执行,但高度取决于注射器的类型和流速。但是,设备在加压时可能不足,而在减压时可能过量,这表明在这些阶段可能需要保持警惕并适当调整剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号