Cyclic torsional loading tests were carried out in the laboratory using various implant systems, in order to clarify differences between the systems in loosening of abutment screws. Six samples from six commercially available abutment systems were used, giving a total of 36 samples. Four of the systems used internal connections, and two used external connections. The abutment screw for each system was tightened to a torque value specified by the manufacturer, and after 5 min, the loosening torque was measured using a digital torque meter. Measurements were taken twice, and a second measurement was taken as a reference value. A cyclic torsional loading test with 100,000 cycles was performed on the sample, and the loosening torque was again measured after the test. In conclusion, loosening of the abutment screw occurred as a result of cyclic torsional loading, and the degree of loosening varied with each implant system.
机译:在实验室中使用各种植入系统进行了周期性扭转载荷测试,目的是弄清系统之间在松开基台螺丝方面的差异。使用了来自六个市售基台系统的六个样本,总共提供了36个样本。其中四个系统使用内部连接,两个使用外部连接。将每个系统的基台螺丝拧紧到制造商指定的扭矩值,并在5分钟后,使用数字扭矩计测量松动扭矩。进行两次测量,并将第二次测量作为参考值。对样品进行了100,000次循环的循环扭转载荷测试,测试后再次测量了松动扭矩。总之,由于周期性的扭转载荷,基台螺丝发生松动,并且松动程度随每个植入系统而变化。