首页> 外文期刊>Hong Kong journal of emergency medicine. >Case Series of Inflation Pressures of the Endotracheal Tube Cuffs among Intubated Patients at an Accident and Emergency Department in Hong Kong
【24h】

Case Series of Inflation Pressures of the Endotracheal Tube Cuffs among Intubated Patients at an Accident and Emergency Department in Hong Kong

机译:香港急症室插管患者气管插管充气压力的病例系列

获取原文
           

摘要

Despite the frequent use of endotracheal tubes in daily practice of emergency medicine, not much emphasis has been put on the measurement of cuff pressure. We attempted to conduct a case series study of prospectively recruited patients to illustrate the risk of inappropriate cuff inflation without manometry. It was found that the pilot balloon palpation was an inaccurate method to estimate cuff under- or over-inflation, with large discrepancies when compared with an objective measure. Such discrepancies might be independent of doctors' and nurses' working experience in airway management. Use of an aneroid manometer remains an economical and cost-effective way to obtain an instantaneous yet reasonably reliable result and should become a routine in emergency departments.
机译:尽管在急诊医学的日常实践中经常使用气管插管,但是对袖带压的测量没有太多的重视。我们试图对预期招募的患者进行案例研究,以说明在没有测压的情况下不适当的袖带充气的风险。结果发现,飞行员触诊球囊触诊是估计袖带充气不足或过度充气的不准确方法,与客观测量相比差异较大。这样的差异可能独立于医生和护士在气道管理方面的工作经验。使用无液压力计仍然是获得瞬时但合理可靠的结果的经济且经济有效的方法,并且应成为急诊部门的常规操作。