首页> 外文期刊>Health >Exploring the useful exposure range limits of three intraoral image receptors for various tube potential, tube current and exposure time settings
【24h】

Exploring the useful exposure range limits of three intraoral image receptors for various tube potential, tube current and exposure time settings

机译:探索三种口腔内图像接收器对于各种管电势,管电流和曝光时间设置的有用曝光范围限制

获取原文
           

摘要

Objectives: To determine the useful exposure range limits of three intraoral image receptors of different technology when exposed to different X-ray beam spectra, dose and dose rate levels. Study Design: A dental X-ray unit offering a wide range of tube potential, tube current and exposure time settings was used to expose a dental quality control phantom. The receptors that were used to capture the radiographic images of the phantom were: the Kodak Insight, the Kodak RVG-6000 and the Duerr Vistascan system. The images that were produced over a wide range of exposure factor settings were evaluated in terms of diagnostic quality by three experienced radiologists. Results: The number of images with acceptable diagnostic quality was in total 1257; 310 with Insight, 331 with RVG 6000 and 616 with Vistascan. At 60 kV, diagnosable images were produced with doses ranging from 0.44-1.56 mGy for the Insight film 0.44-2.82 mGy for the RVG 6000 and 0.22-4.93 mGy for the Vistascan system. At 70 kV, the respective ranges were 0.39-1.28 mGy for the Insight film 0.31-2.55 mGy for the RVG6000 and 0.30-3.46 mGy for the Vistascan system. Conclusions: The Vistascan exhibited the widest useful exposure range and required the least exposure to produce a diagnosable image at almost all tube potential settings. The RVG 6000 exhibited a slightly wider useful exposure range than the Insight film, with almost the same dose requirements especially in higher Kv settings.
机译:目的:确定三种技术不同的口腔内图像接收器在不同的X射线束光谱,剂量和剂量率水平下的有效照射范围极限。研究设计:牙科X射线设备提供了广泛的管电势,管电流和曝光时间设置,用于曝光牙科质量控制体模。用于捕获体模射线照相图像的接收器是:柯达Insight,柯达RVG-6000和Duerr Vistascan系统。由三位经验丰富的放射科医生根据诊断质量评估了在各种曝光因子设置下产生的图像。结果:可以接受的诊断质量图像总数为1257;结果为255。配备Insight的310,配备RVG 6000的331和配备Vistascan的616。在60 kV电压下,可产生可诊断的图像,对于Insight膜,剂量范围为0.44-1.56 mGy(对于RVG 6000为0.44-2.82 mGy),对于Vistascan系统为0.22-4.93 mGy。在70 kV时,Insight膜的相应范围为0.39-1.28 mGy,RVG6000为0.31-2.55 mGy,而Vistascan系统为0.30-3.46 mGy。结论:Vistascan展示了最大的有用曝光范围,并且在几乎所有显像管电势设置下需要最少的曝光即可产生可诊断的图像。 RVG 6000的有效曝光范围比Insight胶片略宽,剂量要求几乎相同,尤其是在较高Kv设置下。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号