首页> 外文期刊>Head Face Medicine >Clinical management and microscopic characterisation of fatique-induced failure of a dental implant. Case report
【24h】

Clinical management and microscopic characterisation of fatique-induced failure of a dental implant. Case report

机译:疲劳引起的牙科植入物故障的临床管理和微观表征。案例报告

获取原文
           

摘要

Background Osseointegrated endosseous implants are widely used for the rehabilitation of completely and partially edentulous patients, being the final prosthodontic treatment more predictable and the failures extremely infrequent. A case of fracture of an endosseous dental implant, replacing the maxillary first molar, occurring in a middle-age woman, 5 years after placement is reported. Materials and methods The difficult management of this rare complication of implant dentistry together with the following rehabilitation is described. Additionally, the authors performed an accurate analysis of the removed fractured implant both by the stereomicroscope and by the confocal laser scanning microscope. Results and discussion The fractured impant showed the typical signs of a fatigue-induced fracture in the coronal portion of the implant together with numerous micro-fractures in the apical one. Three dimensional imaging performed by confocal laser scanning microscope led easily to a diagnosis of "fatigue fracture" of the implant. The biomechanical mechanism of implant fractures when overstress of the implant components due to bending overload is discussed. Conclusion When a fatigue-induced fracture of an dental implant occurs in presence of bending overload, the whole implant suffers a deformation that is confirmed by the alterations (micro-fractures) of the implant observable also in the osseointegrated portion that is easily appraisable by the use of stereomicroscope and confocal laser scanning microscope without preparation of the sample.
机译:背景骨整合式骨内植入物被广泛用于完全和部分缺牙的患者的康复,这是最终的修复牙齿治疗,更可预测,且失败极少发生。据报道,在中年妇女中,放置5年后发生了骨内植入物骨折,取代上颌第一磨牙的骨折。材料和方法描述了这种罕见的种植牙并发症的难以处理以及随后的康复治疗。此外,作者通过立体显微镜和共聚焦激光扫描显微镜对取出的骨折植入物进行了准确的分析。结果与讨论断裂的植入物显示出植入物冠状部分疲劳诱发的骨折的典型体征,而在根尖则显示出许多微骨折。通过共聚焦激光扫描显微镜进行的三维成像很容易诊断出植入物的“疲劳断裂”。讨论了由于弯曲过载导致的植入物组件过应力时植入物骨折的生物力学机制。结论当在弯曲过载的情况下发生牙科植入物的疲劳引起的骨折时,整个植入物都会发生变形,这种变形可以通过植入物的整合(在骨整合部分中观察到)的变化(微骨折)来证实,这种变形很容易被评估。使用体视显微镜和共聚焦激光扫描显微镜无需准备样品。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号