Late dissociation of the polyethylene liner from a modular acetabular metal shell after primary total hip arthroplasty: a report of five cases

机译:初次全髋关节置换术后聚乙烯内衬从模块化髋臼金属壳中的晚期分离:报告5例

获取原文
           

摘要

Modular designs of hip prostheses have become popular recently. Along with complications inherent in all hip arthroplasty systems, modular systems have the additional potential for dissociation of components. Five male patients underwent total hip arthroplasties, in which all of the acetabular components were Harris-Galante II porous acetabular cups. Many years after the operation, the polyethylene liners were dissociated without any previous trauma or dislocation of the femoral heads, these dissociations and dislodgements were managed with open reduction. This complication can be predicted from clinical symptoms and signs. Roentgenograms must be taken and carefully compared to previous roentgenograms. We postulated two causes for the dissociation. First, the polyethylene liner was not fixed securely within the acetabular metal shell at the time of operation. Second, the locking mechanism of the acetabular metal shell was not strong enough to firmly hold the polyethylene liner within the acetabular metal shell. It does warrant that certain precautions must be taken when implanting modular components. The locking mechanism of the harris-Galante II porous acetabular component is mechanically weak and fails easily, therefore its design must be improved in an attempt to prevent postoperative dissociation of the polyethylene liner.
机译:髋关节假体的模块化设计最近变得很流行。除所有髋关节置换系统固有的并发症外,模块化系统还具有分离组件的额外潜力。五名男性患者接受了全髋关节置换术,其中所有髋臼成分均为Harris-Galante II多孔髋臼杯。手术后很多年,聚乙烯衬里被分离,没有任何先前的创伤或股骨头脱位,这些分离和移位均通过切开复位处理。可以从临床症状和体征预测这种并发症。必须拍摄X线照片,并与以前的X线照片进行仔细比较。我们推测了两种分离的原因。首先,在操作时,聚乙烯内衬未牢固地固定在髋臼金属壳内。其次,髋臼金属壳的锁定机制不够牢固,无法将聚乙烯内衬牢固地固定在髋臼金属壳内。它确实保证在植入模块化组件时必须采取某些预防措施。 harris-Galante II多孔髋臼组件的锁定机制在机械上较弱并且很容易失效,因此必须改进其设计,以防止聚乙烯衬垫的术后分离。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号