首页> 外文期刊>Hand >Sensitivity and Specificity of Skyline and Carpal Shoot-Through Fluoroscopic Views of Volar Plate Fixation of the Distal Radius: A Cadaveric Investigation of Dorsal Cortex Screw Penetration
【24h】

Sensitivity and Specificity of Skyline and Carpal Shoot-Through Fluoroscopic Views of Volar Plate Fixation of the Distal Radius: A Cadaveric Investigation of Dorsal Cortex Screw Penetration

机译:Sky骨远侧掌侧板固定术的天际线和腕骨穿透荧光透视图的敏感性和特异性:背皮质螺钉穿透的尸体调查

获取原文
           

摘要

Background: Volar plate fixation of distal radius fractures can result in soft tissue injuries. Abnormal contour of the dorsal cortex of the distal radius provides difficulties in discerning screw penetration on standard radiographs. The skyline and carpal shoot-through views are additional views to improve dorsal cortex visibility. We report on the sensitivity and specificity of determining screw protrusion with these views. Methods: Seven fresh frozen cadavers were instrumented with a distal radius volar locking plate. Initial screw length was determined by depth gauge measurement. A dorsal dissection of the wrist was performed to detect screw penetration. Protruding screws were documented and replaced with screws of the appropriate length and deemed as baseline. Screws were then sequentially lengthened by 2 and 4 mm. Skyline and carpal shoot-through views were obtained at baseline, 2 mm, and 4 mm. The images were randomized and compiled into an untimed survey asking orthopedic surgeons to determine whether screws were penetrating through the dorsal cortex. Results: Based on depth gauge measurements, 4 out of 44 (9.1%) volar plate locking screws penetrated the dorsal cortex, as confirmed with dorsal dissection. Sensitivities for the skyline and carpal shoot-through views were 75% and 86% ( P ≤ .001), respectively, for 2-mm protrusions, and 76% and 89% ( P ≤ .001), respectively, for 4-mm screw protrusions. Specificities were 85% and 84% for the skyline and shoot-through views, respectively. Conclusions: We believe that the carpal shoot-through view has utility and can be implemented to augment standard intraoperative views, and may decrease the incidence of screw protrusion resulting in soft tissue injuries.
机译:背景:Vol骨远端of骨钢板固定可能导致软组织损伤。远端radius骨背皮质的轮廓异常,很难在标准X线片上辨别螺钉的穿透力。天际线和腕骨直通视图是可改善背皮质可见性的其他视图。我们用这些观点报告了确定螺钉突出的敏感性和特异性。方法:对7只新鲜冷冻的尸体进行radius骨远端掌侧锁定板的检测。初始螺钉长度通过深度计测量确定。进行腕的背侧解剖以检测螺钉的穿透。记录突出的螺钉,并用适当长度的螺钉代替,并视为基线。然后将螺丝依次加长2和4毫米。在基线,2毫米和4毫米处获得天际线和腕骨穿透视图。这些图像被随机化,并编入一个不定时的调查中,要求整形外科医生确定螺钉是否穿过背皮质。结果:根据深度计的测量结果,在44颗(9.1%)掌骨固定螺钉中,有4颗穿透了背侧皮质,经背侧解剖证实。对于2毫米突出物,天际线和腕部穿透视图的敏感度分别为75%和86%(P≤.001),对于4毫米突起,其敏感性分别为76%和89%(P≤.001)。螺丝突起。天际线和直射景物的特异性分别为85%和84%。结论:我们认为腕骨穿通视野是有用的,可以用来扩大标准的术中视野,并且可以减少螺钉突出的发生率,从而导致软组织损伤。

著录项