首页> 外文期刊>The Keio Journal of Medicine >Correlation of the Blood Test Results Obtained between Assays Using Microliter-scale Fingertip Blood Samples Collected with a Novel Blood Collection Device and Conventional Venous Blood Assays: a Secondary Publication in English
【24h】

Correlation of the Blood Test Results Obtained between Assays Using Microliter-scale Fingertip Blood Samples Collected with a Novel Blood Collection Device and Conventional Venous Blood Assays: a Secondary Publication in English

机译:使用新型血液收集装置收集的微升级指尖血样与常规静脉血液测定之间的血液测试结果之间的相关性:英语的二次出版

获取原文
           

摘要

In this study, we investigated the correlations between biochemical and hematological test results obtained using microliter-scale fingertip blood samples collected with a newly developed blood collection device and those obtained using conventional venous blood. Eighty volunteer subjects were enrolled in this study. Blood samples were drawn from the fingertip of the ring finger by a single puncture, and 60-μL samples were promptly and accurately aspirated into a blood collection chip. Then the chip was tightly sealed in a chip container and was shaken to mix the contents without dispersion. For biochemical tests other than that for HbA1c, blood was collected without anticoagulant and centrifuged to obtain 15?μL of serum which was then diluted with 190?μL of physiological saline for the assay. For hematological tests and the test for HbA1c, the sample was assayed with blood collected using EDTA-2?K. Good correlations were obtained between the test results of the assay using fingertip blood and that using venous blood. The correlation coefficients were ≥0.97 for TG, T-CHO, HDL-C, LDL-C, GLU, ALT, γ-GTP, UA, BUN, and HbA1c and ≥0.95 for WBC, RBC,?Hgb, and Hct. These results suggest that our microliter-scale blood testing system is comparable to assays using venous blood and may be useful as a rapid and simple test to determine basic clinical parameters that are close to the reference intervals.
机译:在这项研究中,我们调查了使用新开发的采血设备采集的微升规模指尖血液样品获得的生化和血液学测试结果与使用常规静脉血获得的结果之间的相关性。这项研究招募了80名志愿者。只需一次穿刺就从无名指的指尖抽取血样,然后将60μL的样本迅速准确地吸到采血芯片中。然后将切屑紧密密封在切屑容器中,并摇动以混合内容物而不会分散。对于除HbA1c以外的生化测试,在不使用抗凝剂的情况下收集血液并离心以获得15?L的血清,然后将其用190?L的生理盐水稀释以进行测定。对于血液学检验和HbA1c检验,用EDTA-2?K采集的血液对样品进行分析。使用指尖血液和静脉血的检测结果之间具有良好的相关性。 TG,T-CHO,HDL-C,LDL-C,GLU,ALT,γ-GTP,UA,BUN和HbA1c的相关系数≥0.97,而WBC,RBC,ΔHgb和Hct的相关系数≥0.95。这些结果表明,我们的微升规模的血液测试系统可与使用静脉血的分析相媲美,并且可以作为一种快速简单的测试来确定接近参考区间的基本临床参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号