首页> 外文期刊>Texas Heart Institute journal / >Use of ViperSlide Lubricant to Extract Entrapped Sheath after Severe Radial Artery Spasm during Coronary Angiography
【24h】

Use of ViperSlide Lubricant to Extract Entrapped Sheath after Severe Radial Artery Spasm during Coronary Angiography

机译:在冠状动脉造影期间严重Rad动脉痉挛后使用ViperSlide润滑剂提取被困的鞘

获取原文
           

摘要

Radial artery spasm is a known complication of transradial cardiac catheterization. However, severe spasm with sheath entrapment is rare. We describe such a case, and the condition's response to an alternative removal method after conventional efforts failed. A 68-year-old man presented for coronary angiography. We introduced a 5F sheath into the right radial artery, but, because of severe arterial spasm, we could not aspirate blood from the sheath or retract it. We sedated the patient and waited for the spasm to subside; however, the radial sheath remained entrapped. Nitroglycerin injection enabled blood aspiration and vasodilator injection, but not sheath removal. Finally, we injected ViperSlide lubricant into the sheath for its rapid, easy extraction. When sedation and vasodilator therapy fail, we recommend using ViperSlide for radial sheath removal before applying nerve block or general anesthesia.
机译:ial动脉痉挛是trans动脉心脏导管插入术的已知并发症。然而,严重的痉挛伴鞘管夹带很少见。我们描述了这种情况,并且在常规工作失败后,条件对其他替代方法的响应。一名68岁男子接受了冠状动脉造影检查。我们在右radial动脉中插入了5F护套,但是由于严重的动脉痉挛,我们无法从护套中抽出血液或将其缩回。我们为病人镇静,等待痉挛消退。但是,sheath骨鞘仍然被夹住。硝酸甘油注射可以进行血液抽吸和血管扩张剂注射,但不能去除鞘。最后,我们将ViperSlide润滑剂注入护套中,以快速,轻松地提取。当镇静和血管扩张药治疗失败时,我们建议在应用神经阻滞或全身麻醉之前,使用ViperSlide去除radial骨鞘。