Sensitivity Study to Evaluate the Dosimetric Impact of Off-Axis Ratio Profiles Misalignment on TomoTherapy Second Dose Validation

机译:评估偏轴比轮廓未对准对tomotherapy第二剂量验证的剂量学影响的敏感性研究

获取原文
获取外文期刊封面目录资料

摘要

Accurate dose planning and delivery are very important in the intensity modulated radiation therapy. For helical TomoTherapy dose validation, a TomoTherapy second check software, called MU-Tomo, has been developed using archived patient documents, initial coordinates and planned dose of the point of calculation, and common dosimetric functions. Based on this software, sensitivity studies on 50 patient cases have been evaluated to show the impact of off-axis ratio profile misalignment on point dose calculation. Off-axis ratio is defined as the dose profile normalized to its maximum dose value. Sensitivity studies were done for three scenarios: oscillating the fluctuation regions of two off-axis profiles, shifting the profiles, and rotating the profiles. The result of the oscillation trial is linear along the change of longitudinal off-axis ratio (OARy), while oscillating the lateral off-axis ratio (OARx) has little influence on the dose calculation. For shifting, the variation in the percentage difference from the non-shifting value is about 15 times larger in OARy modification than in OARx modification. Rotating OARx by ±6° gave less than 1.5% ± 0.20% difference compared to the non-rotating value. Rotating OARy by ±1° changes the result more than 5% ± 2.69%. Therefore, for helical TomoTherapy dose validation, commissioned OARy profiles are more sensitive than OARx to oscillation, shifting and rotating. As a result, different tolerances for OARx and OARy may be required for annual quality assurance.
机译:在调强放射治疗中,准确的剂量计划和递送非常重要。为了进行螺旋TomoTherapy剂量验证,已经开发了一种称为MU-Tomo的TomoTherapy第二检查软件,它使用已存档的患者文档,初始坐标和计算点的计划剂量以及常用的剂量测定功能。基于该软件,对50例患者的敏感性研究进行了评估,以显示离轴比率曲线未对准对点剂量计算的影响。离轴比定义为归一化为其最大剂量值的剂量曲线。针对以下三种情况进行了灵敏度研究:振荡两个离轴轮廓的波动区域,移动轮廓以及旋转轮廓。振荡试验的结果沿纵向离轴比(OARy)的变化呈线性,而横向离轴比(OARx)的振荡对剂量计算的影响很小。对于移位,与OARx修改相比,OARy修改中与非移动值的百分比差异的变化大约大15倍。与非旋转值相比,将OARx旋转±6°产生的差异小于1.5%±0.20%。将OARy旋转±1°可使结果变化超过5%±2.69%。因此,对于螺旋TomoTherapy剂量验证,已委托的OARy配置文件比OARx对振荡,移动和旋转更敏感。结果,可能需要为OARx和OARy设置不同的公差,以保证年度质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号