首页> 外文期刊>Química Nova >Determina??o potenciométrica em fluxo de cloreto de cetilpiridinio em desinfectantes bucais
【24h】

Determina??o potenciométrica em fluxo de cloreto de cetilpiridinio em desinfectantes bucais

机译:电位法测定漱口液中十六烷基氯化吡啶鎓的流量

获取原文
           

摘要

The work describes a new procedure for cetylpyridinium chloride determination in oral disinfectants, based on a flow-injection system with potentiometric detection. The determination was based on the measurement of picrate concentration decrease as result of ion-pair reaction with the analyte present in the injected sample. In the optimised set-up the sample injection volume was kept at 400 μL and merged downstream with the reagent solution containing 1,0 x10-5 mol/L of picrate adjusted to pH 5.0 with citrate/citric acid buffer. The flow rate was fixed at 8 mL/min and the reactor length at 40 cm. The proposed procedure enables the determination of cetylpyridinium in the analytical range of 5,0x10-6 - 7,5x10-5 mol/L at a sampling rate of 60/h. The results for real samples had a precision better than 3% and were comparable to the labelled values.
机译:这项工作介绍了一种基于带电位检测功能的流动注射系统,用于测定口腔消毒剂中十六烷基氯化吡啶的新方法。该确定是基于与进样样品中存在的分析物发生离子对反应而导致的苦味酸盐浓度降低的测量结果。在优化设置中,样品进样量保持在400μL,并与下游的试剂溶液合并,该溶液含有1,0x10-5 mol / L苦味酸盐,并用柠檬酸盐/柠檬酸缓冲液调节至pH 5.0。流速固定为8 mL / min,反应器长度为40 cm。所提出的方法能够以60 / h的采样率在5,0x10-6-7,5x10-5 mol / L的分析范围内测定鲸蜡基吡啶。真实样品的结果精度优于3%,与标记值相当。