首页> 外文期刊>Practical Laboratory Medicine >Assessment of liver function tests on Piccolo Xpress point of care chemistry analyzer in a pediatric hospital
【24h】

Assessment of liver function tests on Piccolo Xpress point of care chemistry analyzer in a pediatric hospital

机译:儿科医院用Piccolo Xpress护理化学分析仪进行肝功能测试的评估

获取原文
       

摘要

Objectives Point of care testing (POCT) contributes to diagnosis and monitoring with fast testing time and easily performed assays. We evaluated the Abaxis Piccolo Xpress point of care chemistry analyzer using the Liver Panel Plus discs for our pediatric patient population at Texas Children's Hospital. Design and methods Analytical performance was evaluated for precision and linearity using quality control materials and commercially available verification samples. Comparison studies were performed between Piccolo Xpress analyzer and Vitros 5600 analyzer using patient samples. Interference studies were carried out using nine different patient pool sera. Lipemia interference was removed using LipoClear for severely lipemic sample pools. Results Precision of all tests was excellent (CVs<5% for all measured analytes except TBIL). All assays were linear and accurate within the allowable total error. Comparison studies showed that three analytes (amylase, GGT and TBIL) had statistically significant bias. Interference study results did not exceed the total allowable error for hemoglobin (<150mg/dL), bilirubin (<15mg/dL) and lipemia (<400mg/dL except ALT, GGT and TP). LipoClear treatment removed lipemia interference for all analytes except total protein. Conclusions The Piccolo Xpress chemistry analyzer showed an acceptable analytical performance for precision, linearity and interference from common substances. Increased bias for three analytes in comparison studies could be due to different methodologies. Highlights ? Liver Panel Plus analytes were evaluated on Piccolo Xpress chemistry analyzer. ? All analytes except bilirubin, amylase, and GGT showed acceptable performance for precision and interference in our pediatric population.
机译:目的即时检验(POCT)有助于通过快速的检验时间和易于执行的化验来进行诊断和监控。我们使用Liver Panel Plus光盘评估了Abaxis Piccolo Xpress护理化学分析仪,用于德克萨斯儿童医院的儿科患者人群。设计和方法使用质量控制材料和市售验证样品评估分析性能的准确性和线性。使用患者样本在Piccolo Xpress分析仪和Vitros 5600分析仪之间进行了比较研究。使用九种不同的患者库血清进行干扰研究。使用LipoClear消除了严重血脂样本池的血脂干扰。结果所有测试的精度都非常出色(除TBIL之外,所有测得的分析物的CVs <5%)。所有测定均线性且在允许的总误差内准确。比较研究表明,三种分析物(淀粉酶,GGT和TBIL)具有统计学上的显着偏差。干扰研究结果未超过血红蛋白(<150mg / dL),胆红素(<15mg / dL)和血脂(<400mg / dL,但ALT,GGT和TP除外)的总允许误差。 LipoClear处理消除了除总蛋白以外的所有分析物的血脂干扰。结论Piccolo Xpress化学分析仪在精密度,线性和常见物质干扰方面显示出可接受的分析性能。比较研究中三种分析物的偏倚增加可能是由于不同的方法。强调 ?在Piccolo Xpress化学分析仪上评估Liver Panel Plus分析物。 ?除胆红素,淀粉酶和GGT以外的所有分析物在我们的儿科人群中均显示出可接受的精确度和干扰性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号