Improved assay protocol for measurement of ammonia on the Roche Cobas 8000 automated platform

机译:在Roche Cobas 8000自动化平台上改进的用于测量氨的测定规程

获取原文
       

摘要

IntroductionAmmonia is a metabolite of protein catabolism that, when elevated, may be toxic for tissues, especially for the central nervous system. Elevated ammonia in blood is an indicator and a prognostic factor for hepatic and kidney disease or inherited metabolic disorders in nitrogen metabolism. The accuracy of ammonia determination is influenced by sampling condition, handling, storage and assay itself. Our and other laboratories have been experiencing high frequencies sample error flags while measuring ammonia with glutamate dehydrogenase method on Roche Cobas 8000 platform. To reduce the number of error flags we adapted Roche NH3L protocol by incorporation of an additional onboard routine step for sample pre-dilution.Material and methodsThe AMC NH3L is an adaptation of Roche protocol that uses four fold pre-dilution of the sample in the rerun prior to the analysis. It was assessed for 1.occurrence of absorbance error flags, 2.precision, 3.correlation with Roche method and 4.interference by hemolysis, icterus and lipemia.ResultsThe AMC NH3L adaptation demonstrates acceptable within-run and total precision. Comparison studies show no differences between the Roche rerun application and AMC NH3L adaptation. The AMC NH3L adaptation solves 78% of absorbance errors and for samples with high ammonia concentration is less affected by interferences from icterus and hemolysis than the Roche rerun application.ConclusionThe AMC NH3L adaptation is less prone to instrument error flags and for samples with high ammonia concentration, is more robust to endogenous interferences. The AMC NH3L adaptation is viable alternative to the Roche protocol for the ammonia measurement.
机译:简介氨是蛋白质分解代谢的代谢产物,当升高时,对组织,特别是对中枢神经系统可能有毒。血液中氨含量升高是肝肾疾病或氮代谢中遗传性代谢异常的指标和预后因素。氨气测定的准确性受采样条件,处理,存储和测定本身的影响。我们的实验室和其他实验室在使用Roche Cobas 8000平台上的谷氨酸脱氢酶方法测量氨气时一直遇到高频样品错误标志。为了减少错误标志的数量,我们通过结合额外的机上常规步骤进行样品预稀释来适应Roche NH3L协议。材料和方法AMC NH3L是Roche协议的改编,在重新运行中使用了四倍的样品预稀释在分析之前。评估了1.吸光度错误标志的出现,2。精密度,3。与罗氏方法的相关性以及4.溶血,黄疸和脂血的干扰。比较研究表明,Roche重播应用程序与AMC NH3L适应程序之间没有差异。与Roche rerun应用相比,AMC NH3L适应性解决了78%的吸光度误差,对于高氨浓度的样品,不受黄疸和溶血干扰的影响比Roche rerun应用更少。结论AMC NH3L适应性的仪器不易出现仪器错误标志以及对于高氨浓度的样品对内源性干扰更强大。对于氨气测量,AMC NH3L适应方案是罗氏协议的可行替代方案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号