Absolute distance measurement by means of a self-mixing interferometer (SMI) can be obtained by modulating the laser frequency. This modulation is typically achieved through the modulation of the laser injection current with a triangle waveform. Various strategies have been proposed to increase the performances and recent researches have shown real-time performance of SMI with resolution reaching 100 ??m for distances up to 2 m. In the present paper, we demonstrate for the first time, both experimentally and by modeling, that with high coupling factors between the laser and the target, disappearance of interferometric fringes occurs that can strongly affect the measurement reliability.
机译:借助于自混合干涉仪(SMI),可以通过调制激光频率来获得绝对距离测量。通常通过以三角波形调制激光注入电流来实现该调制。已经提出了各种策略来提高性能,并且最近的研究表明,对于2M的距离,SMI的实时性能可以达到100?m的分辨率。在本文中,我们首次通过实验和模型证明,在激光与目标之间具有高耦合因子的情况下,会出现干涉条纹消失,这会严重影响测量的可靠性。