Modification of conventional radiography systems into a digital radiography system was conducted at the Laboratory of Physics Unnes . The image quality of digital radiography system that was developed in the Laboratory of Medical Physics Unnes need to be tested. The computed radiography (CR) system owned by Diponegoro National Hospital was chosen as a comparison with contrass to noise ratio (CNR) methode. Radiograph image generated by both radiography system then analyzed by software based on MATLAB. This study uses a copper plate phantom exposed to X-rays with a variation of anode-cathode voltage of 50 kV, 60 kV, 70 kV and 80 kV. For each voltage to vary the electrical current and expotion time of 1.6 mAs, 2.0 mAs, mAs 4.0, 8 mAs, 16 mAs and 32 mAs. Comparison of CNR for both systems to variations of tube voltage (kV) have a fixed value for the current and expotion time 4 mAs. Comparison of CNR for both systems to variations in current and ekspotion time (mAs) selected tube voltage 70kV.
机译:在联合国大学物理实验室进行了将常规放射线照相系统改造成数字放射线照相系统的工作。需要对在Unnes的医学物理实验室开发的数字X射线照相系统的图像质量进行测试。与敌对噪声比(CNR)方法进行比较,选择了迪波内哥罗国立医院拥有的计算机放射成像(CR)系统。两种射线照相系统生成的射线照相图像,然后通过基于MATLAB的软件进行分析。这项研究使用的铜板体模暴露于X射线,阳极-阴极电压的变化为50 kV,60 kV,70 kV和80 kV。对于每个电压,电流和曝光时间分别为1.6 mA,2.0 mA,4.0、8 mA,16 mA和32 mA。两种系统的CNR与管电压(kV)变化的比较对于电流和曝光时间4 mAs具有固定值。比较两种系统的CNR与选定的70kV管电压的电流和吸波时间(mAs)的变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号