首页> 外文期刊>Romanian reports in physics >Long term stability of the performance of a clinical linear accelerator and Z-score assessment for absorbed dose to water quantity
【24h】

Long term stability of the performance of a clinical linear accelerator and Z-score assessment for absorbed dose to water quantity

机译:临床线性加速器性能的长期稳定性和吸收剂量对水量的Z评分评估

获取原文
           

摘要

Assessment of the z-score was used to evaluate the long term stability of the performance of a clinical linear accelerator. The absorbed dose to water in high energy photon beam was determined at the linear accelerator owned by the clinical dosimetry laboratory in the radiotherapy department of the "Coltea" Clinical Hospital. The z-score was calculated for the absorbed dose measurements made at reference conditions according to IAEA-398 standard, in 2011 and in 2015 following repairs made to the linear accelerator. The measurements were conducted using water proof ionization chambers in a water equivalent phantom at the reference point for a 10 cm × 10 cm field size and at SSD = 1 m. For each ionization chamber the reference points, were determined according to radiation beam quality. The measurements were performed in a photon beam at reference points of 5 g cm-2 and 10 g cm-2. z-score assessment for absorbed dose to water quantity was evaluated according to ISO 13528:2005. The values obtained were found to be in the limits of statistical performance, z = [-2; 2], concluding that laboratory performance is satisfactory.
机译:z分数的评估用于评估临床线性加速器性能的长期稳定性。高能光子束对水的吸收剂量由“ Coltea”临床医院放射治疗部门的临床剂量测定实验室拥有的线性加速器确定。在2011年和2015年对线性加速器进行维修后,根据IAEA-398标准在参考条件下进行的吸收剂量测量计算了z分数。使用防水电离室在相当于水的体模中在参考点上进行10 cm×10 cm的场大小并在SSD = 1 m时进行测量。对于每个电离室,根据辐射束质量确定参考点。在参考点为5 g cm-2和10 g cm-2的光子束中进行测量。根据ISO 13528:2005对吸收剂量对水量的z评分进行评估。发现获得的值在统计性能的极限内,z = [-2; 2],认为实验室的性能令人满意。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号