Analysis of damping characteristics of arterial catheter blood pressure monitoring in a large intensive care unit

机译:大型重症监护室中动脉导管血压监测的阻尼特性分析

获取原文
       

摘要

BACKGROUND. For many reasons, the invasive measurement of systolic and diastolic blood pressure should be accurate. Accuracy is determined, in part, by the damping characteristics of the arterial catheter blood pressure monitoring system. OBJECTIVES. To ascertain the damping characteristics of arterial catheter blood pressure monitoring in a large tertiary intensive care unit (ICU) and to elicit any causes of under- or over-damping of the measurement systems. METHODS. A cross-sectional, observational study of arterial line measurements in a large general ICU. The coefficient of damping (CoD) was calculated from the waveform generated from a 'fast flush'. RESULTS. Thirty systems (19%) were adequately damped (CoD 0.4 - 0.8), 56 (37%) were overdamped, and 68 (44%) were underdamped. We did not find that poor damping characteristics were associated with the age of the arterial catheter or the type of catheter used. CONCLUSION. Most systems observed in this study were inappropriately damped, which would result in the inaccurate display of the waveform and systolic and diastolic pressures.
机译:背景。由于许多原因,收缩压和舒张压的侵入性测量应准确。准确性部分取决于动脉导管血压监测系统的阻尼特性。目标为了确定大型三级重症监护病房(ICU)的动脉导管血压监测的阻尼特性,并找出导致测量系统阻尼过低或过高的任何原因。方法。大型普通ICU中动脉线测量的横断面观察研究。阻尼系数(CoD)由“快速冲洗”生成的波形计算得出。结果。三十个系统(19%)被充分阻尼(CoD 0.4-0.8),56(37%)被过度阻尼,68(44%)被削弱。我们没有发现不良的阻尼特性与动脉导管的年龄或所用导管的类型有关。结论。在这项研究中观察到的大多数系统都不适当地阻尼,这将导致波形,收缩压和舒张压显示不正确。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号