首页> 外文期刊>Sage Open Medicine >Air bubbles are released by thoracic endograft deployment: An in vitro experimental study:
【24h】

Air bubbles are released by thoracic endograft deployment: An in vitro experimental study:

机译:气泡通过胸腔内植入物释放:一项体外实验研究:

获取原文
           

摘要

Purpose:Embolic stroke is a dreaded complication of thoracic endovascular aortic repair. The prevailing theory about its cause is that particulate debris from atherosclerotic lesions in the aortic wall are dislodged by endovascular instruments and embolize to the brain. An alternative source of embolism might be air trapped in the endograft delivery system. The aim of this experimental study was to determine whether air is released during deployment of a thoracic endograft.Methods:In an experimental benchtop study, eight thoracic endografts (five Medtronic Valiant Thoracic and three Gore TAG) were deployed in a water-filled transparent container drained from air. Endografts were prepared and deployed according to their instructions for use. Deployment was filmed and the volume of air released was collected and measured in a calibrated syringe.Results:Air was released from all the endografts examined. Air volumes ranged from 0.1 to 0.3?mL for Medtronic Valiant Thoracic and from 0.025 to 0.04?mL for Gore T...
机译:目的:栓塞性中风是胸腔内血管主动脉修复的可怕并发症。关于其原因的流行理论是,主动脉壁动脉粥样硬化病变的颗粒碎片被血管内器械清除并栓塞到大脑。栓塞的另一种来源可能是将空气滞留在植入物内的输送系统中。该实验研究的目的是确定在部署胸腔内移植物时是否释放空气。方法:在一项实验性台式研究中,将8例胸腔内移植物(5个Medtronic Valiant Thoracic和3个Gore TAG)放置在一个充满水的透明容器中排干空气。根据其使用说明准备并部署内移植物。记录下展开物,收集释放的空气量并在校准的注射器中进行测量。结果:从所有检查的内移植物中释放出空气。 Medtronic Valiant胸腔的风量范围从0.1到0.3?mL,而Gore T的风量范围从<0.025到0.04?mL。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号