首页> 外文期刊>Revista Brasileira de Medicina do Esporte >Estimativa da composi??o corporal e análise de concordancia entre analisadores de impedancia bioelétrica bipolar e tetrapolar
【24h】

Estimativa da composi??o corporal e análise de concordancia entre analisadores de impedancia bioelétrica bipolar e tetrapolar

机译:双极和四极生物电阻抗分析仪的人体成分估计和一致性分析

获取原文
           

摘要

The aim of this study was to analyze the body composition estimation and agreement of the measurements produced from two bioelectrical impedance analysers in young adult men. Twenty-two apparently healthy subjects (23.6 ± 2.9 yrs; 73.0 ± 6.8 kg; 177.6 ± 6.0 cm; 23.1 ± 1.9 kg/m2), were submitted to bioelectrical impedance measurements by tetra-polar whole body (TWB, Biodynamic Body Composition Analyser, model 310) and bipolar for upper body (BUB, OMRONTM, model HBF-306BL) after 12 h fast. Relative body fat (%BF), lean body mass (LBM) and fat mass (FM) estimation were established from information supplied by two analysers. No significant difference was verified for %BF (10.9 ±4.3 vs. 10.0 ±4.2%; P = 0.09), FM (8.1 ±3.5 vs. 7.4 ±3.5 kg; P = 0.10) or LBM (64.9 ±5.2 vs. 65.5 ±4.9 kg; P = 0.09) between TWB and BUB, respectively. Positive significant correlations (P < 0.001) were found between both analysers for LBM (r = 0.948), FM (r = 0.891) and %BF (r = 0.859). The agreement thresholds were -0.6 ± 3.2 kg for LBM (r = 0.135; P = 0.551), 0.8 ± 4.5% for %BF (r = 0.050; P = 0.824), and 0.6 ± 3.2 kg for FM (r = 0.032; P = 0.886). The results suggest that the body composition estimation by bioelectrical impedance does not seem to be affected by use of different analysers (BUB vs. TWB) or the total body adiposity.
机译:这项研究的目的是分析两个成年男性的生物电阻抗分析仪的身体成分估计值和测量结果的一致性。 22名看似健康的受试者(23.6±2.9岁; 73.0±6.8千克; 177.6±6.0厘米; 23.1±1.9千克/平方米)通过四极全身进行了生物电阻抗测量(TWB,生物动力身体成分分析仪,禁食12小时后,上半身(BUB,OMRONTM,HBF-306BL型)双极。相对脂肪(%BF),瘦体重(LBM)和脂肪质量(FM)的估算是由两个分析人员提供的信息确定的。对于%BF(10.9±4.3对10.0±4.2%; P = 0.09),FM(8.1±3.5对7.4±3.5 kg; P = 0.10)或LBM(64.9±5.2对65.5± TWB和BUB之间分别为4.9千克; P = 0.09)。两种分析仪的LBM(r = 0.948),FM(r = 0.891)和%BF(r = 0.859)之间都存在正显着的相关性(P <0.001)。 LBM的一致性阈值为-0.6±3.2 kg(r = 0.135; P = 0.551),%BF的一致性阈值为0.8±4.5%(r = 0.050; P = 0.824),FM的为0.6±3.2 kg(r = 0.032; P = 0.886)。结果表明,通过生物电阻抗估算身体成分似乎不受使用不同分析仪(BUB与TWB)或全身肥胖的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号