Development of an Automatic Seed Marker Registration Algorithm Using CT and kV X-ray Images

机译:基于CT和kV X射线图像的种子标记自动配准算法的开发

获取原文
           

摘要

PURPOSE: The purpose of this study is to develop a practical method for determining accurate marker positions for prostate cancer radiotherapy using CT images and kV x-ray images obtained from the use of the on-board imager (OBI). MATERIALS AND METHODS: Three gold seed markers were implanted into the reference position inside a prostate gland by a urologist. Multiple digital image processing techniques were used to determine seed marker position and the center-of-mass (COM) technique was employed to determine a representative reference seed marker position. A setup discrepancy can be estimated by comparing a computed COMOBI with the reference COMCT. A proposed algorithm was applied to a seed phantom and to four prostate cancer patients with seed implants treated in our clinic. RESULTS: In the phantom study, the calculated COMCT and COMOBI agreed with COMactual within a millimeter. The algorithm also could localize each seed marker correctly and calculated COMCT and COMOBI for all CT and kV x-ray image sets, respectively. Discrepancies of setup errors between 2D-2D matching results using the OBI application and results using the proposed algorithm were less than one millimeter for each axis. The setup error of each patient was in the range of 0.1+/-2.7~1.8+/-6.6 mm in the AP direction, 0.8+/-1.6~2.0+/-2.7 mm in the SI direction and -0.9+/-1.5~2.8+/-3.0 mm in the lateral direction, even though the setup error was quite patient dependent. CONCLUSION: As it took less than 10 seconds to evaluate a setup discrepancy, it can be helpful to reduce the setup correction time while minimizing subjective factors that may be user dependent. However, the on-line correction process should be integrated into the treatment machine control system for a more reliable procedure.
机译:目的:本研究的目的是开发一种实用的方法,该方法使用CT图像和kV X射线图像(通过使用车载成像器(OBI)获得)来确定前列腺癌放射治疗的准确标记位置。材料与方法:泌尿科医师将三种金种子标记物植入前列腺内的参考位置。使用多种数字图像处理技术确定种子标记位置,并使用质心(COM)技术确定代表性参考种子标记位置。可以通过将计算出的COMOBI与参考COMCT进行比较来估计设置差异。提出的算法已应用于种子体模和四名在我们诊所接受过种子植入的前列腺癌患者。结果:在幻像研究中,计算出的COMCT和COMOBI与COMactual在1毫米内吻合。该算法还可以正确定位每个种子标记,并分别针对所有CT和kV X射线图像集计算COMCT和COMOBI。使用OBI应用程序的2D-2D匹配结果与使用所提出算法的结果之间的设置误差差异,每个轴均小于1毫米。每个患者在AP方向上的设置误差范围为0.1 +/- 2.7〜1.8 +/- 6.6 mm,在SI方向上的设置误差范围为0.8 +/- 1.6〜2.0 +/- 2.7 mm,在-0.9 +/-范围内即使安装误差在很大程度上取决于患者,在横向方向上为1.5〜2.8 +/- 3.0 mm。结论:由于评估设置差异所需的时间少于10秒,因此有助于减少设置校正时间,同时最大程度地减少可能取决于用户的主观因素,这可能会有所帮助。但是,在线校正过程应集成到处理机控制系统中,以实现更可靠的过程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号