High Dose Rate Ir-192 Source Calibration Method with Newly Designed Calibration Jig

机译:采用新设计的校准夹具的高剂量率Ir-192源校准方法

获取原文
           

摘要

Authors have developed highly reproducible calibration method for the Micro-Selection HDR Ir-192 system(Nucletron, Netherland). The new jig has a 10cm radius circular hole in the 30cm x 30cm x 0.2cm acrylic plate, and 5F flexible bronchial tubes are attached around the hole. The source moves along the circle in the tubes an? the ionization chamber is placed vertically at the center of the circular hole(center of the jig). Dose distribution near the center was derived theoretically, and measured with the film dosimetry system. Theoretical calculation and measurement show the error margin below 0.1% for 1mm or 2mm position deviation. We have measured at 12 and 24 points of circle with 1, 6, 11 and 21 second dwell time of source in order to calculate the activity of the source. Measurements have been repeated daily for 50 days. The accuracy and the reproducibility are below 1% error margin. The half life of the source from our measurement is estimated 73.4+/-0.4 days.
机译:作者已经为Micro-Selection HDR Ir-192系统(荷兰的Nucletron)开发了高度可重复的校准方法。新夹具在30cm x 30cm x 0.2cm丙烯酸板上有一个半径为10cm的圆形孔,并且在该孔周围连接了5F柔性支气管。源沿着管中的圆运动吗?电离室垂直放置在圆形孔的中心(夹具的中心)。理论上推导了中心附近的剂量分布,并使用薄膜剂量测定系统进行了测量。理论计算和测量表明,对于1mm或2mm位置偏差,误差容限在0.1%以下。为了计算源的活动性,我们在源的1、6、11和21秒的停留时间对圆的12和24点进行了测量。每天重复测量50天。准确性和可重复性均低于1%的误差范围。根据我们的测量,该光源的半衰期估计为73.4 +/- 0.4天。