BACKGROUND AND OBJECTIVES: Although rare, intraoperative awareness is a severe general anesthesia complication. Anesthesia machine malfunction is one of several causes for anesthetic complications, among them intraoperative awareness. This report aimed at showing a case where the volatile anesthetic monitor has detected machine malfunction which could have led to intraoperative awareness. CASE REPORT: Female patient, 38 years old, 55 kg, physical status ASA I, with right breast cancer, admitted for radical mastectomy and immediate myocutaneous flap reconstruction. Epidural puncture was performed at T8-T9 with a 17-gauge Tuohy epidural needle followed by the introduction of 18-gauge epidural catheter and administration of 0.2% ropivacaine. General anesthesia was then induced, followed by sevoflurane vaporization. Although other monitoring parameters have not detected relevant findings, the inhalational anesthetic monitor has not identified the presence of sevoflurane, thus allowing the diagnostic of vaporizer leakage. CONCLUSIONS: Although rare, intraoperative awareness is a severe complication which should be prevented. Routine and thorough anesthesia equipment inspection before its use may minimize failures. The inhalational anesthetics monitor is useful whenever inhaled anesthetics are being used and may early detect anesthesia machine failures as in this case report.
机译:背景与目的:术中知晓虽然很罕见,但却是一种严重的全身麻醉并发症。麻醉机故障是麻醉并发症的几种原因之一,其中包括术中意识。该报告旨在显示一种情况,其中挥发性麻醉剂监测仪已检测到可能导致术中意识的机器故障。病例报告:女患者,38岁,55公斤,身体状况ASA I,患有右乳腺癌,被接受了根治性乳房切除术和即刻肌皮瓣重建术。在T8-T9时使用17号规的Tuohy硬膜外针进行硬膜外穿刺,然后引入18号规硬膜外导管并施用0.2%罗哌卡因。然后诱导全身麻醉,然后进行七氟醚气化。尽管其他监测参数尚未检测到相关发现,但吸入麻醉监测仪尚未发现七氟醚的存在,因此可以诊断蒸发器泄漏。结论:尽管很少见,但术中意识是严重的并发症,应避免。在使用麻醉设备之前进行常规和彻底的检查可以最大程度地减少故障。每当使用吸入麻醉药时,吸入麻醉药监控器就很有用,并且可以在这种情况下尽早发现麻醉机故障。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号