首页> 外文期刊>Revista Brasileira de Anestesiologia >Dissec??o da parede interna de tubo endotraqueal aramado que causa obstru??o das vias aéreas no intraoperatório sob anestesia geral: relato de caso
【24h】

Dissec??o da parede interna de tubo endotraqueal aramado que causa obstru??o das vias aéreas no intraoperatório sob anestesia geral: relato de caso

机译:全身麻醉下在术中引起气道阻塞的气管导管的内壁解剖:病例报告

获取原文
           

摘要

Endotracheal intubation is performed to establish a secure airway. However, this carries its risks and obstruction of an endotracheal tube (ETT) is a potentially life-threatening event. We report two cases with an obstruction of the resterilized, single use, spiral, reinforced endotracheal tubes by dissection of the internal wall. As a conclusion, we suggest not reusing and resterilizing single tubes in these cases to avoid a complication like dissection of the internal wall of the tube, as this has been the main cause.
机译:进行气管插管以建立安全的气道。但是,这会带来风险,气管插管(ETT)阻塞可能会危及生命。我们报告了两例通过解剖内壁而阻塞了再灭菌的,一次性使用的螺旋形,加强型气管导管。总之,在这种情况下,我们建议不要重复使用和重新消毒单个管子,以免出现诸如解剖管子内壁的并发症,因为这是主要原因。