摘要

Objective The authors have sought to study the calibration of a clinical PKA meter (Diamentor E2) and a calibrator for clinical meters (PDC) in the Laboratory of Ionizing Radiation Metrology at Instituto de Energia e Ambiente - Universidade de São Paulo. Materials and Methods Different qualities of both incident and transmitted beams were utilized in conditions similar to a clinical setting, analyzing the influence from the reference dosimeter, from the distance between meters, from the filtration and from the average beam energy. Calibrations were performed directly against a standard 30 cm3 cylindrical chamber or a parallel-plate monitor chamber, and indirectly against the PDC meter. Results The lowest energy dependence was observed for transmitted beams. The cross calibration between the Diamentor E2 and the PDC meters, and the PDC presented the greatest propagation of uncertainties. Conclusion The calibration coefficient of the PDC meter showed to be more stable with voltage, while the Diamentor E2 calibration coefficient was more variable. On the other hand, the PDC meter presented greater uncertainty in readings (5.0%) than with the use of the monitor chamber (3.5%) as a reference.
机译:目的研究人员试图在圣保罗大学能源学院的电离辐射计量实验室研究临床PKA测量仪(Diamentor E2)和临床测量仪的校准器(PDC)。材料和方法在类似于临床设置的条件下,采用了不同质量的入射光束和透射光束,分析了参考剂量计,仪表之间的距离,过滤和平均光束能量的影响。可以直接在标准30 cm3圆柱室或平行板监测室上进行校准,也可以在PDC仪表上进行间接校准。结果观察到透射光束的能量依赖性最低。 Diamentor E2与PDC仪表以及PDC之间的交叉校准显示了最大的不确定性传播。结论PDC仪表的校准系数随电压变化更稳定,而Diamentor E2的校准系数变化更大。另一方面,与使用监测室(3.5%)作为参考相比,PDC仪表的读数不确定性更高(5.0%)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号