The skin dose tracking system (DTS) that we developed provides a color-coded mapping of the cumulative skin dose distribution on a 3D graphic of the patient during fluoroscopic procedures in real time. The DTS has now been modified to also calculate the kerma area product (KAP) and cumulative air kerma (CAK) for fluoroscopic interventions using data obtained in real-time from the digital bus on a Toshiba Infinix system. KAP is the integral of air kerma over the beam area and is typically measured with a large-area transmission ionization chamber incorporated into the collimator assembly. In this software, KAP is automatically determined for each x-ray pulse as the product of the air kerma/ mAs from a calibration file for the given kVp and beam filtration times the mAs per pulse times the length and width of the beam times a field nonuniformity correction factor. Field nonuniformity is primarily the result of the heel effect and the correction factor was determined from the beam profile measured using radio-chromic film. Dividing the KAP by the beam area at the interventional reference point provides the area-averaged CAK. The KAP and CAK per x-ray pulse are summed after each pulse to obtain the total procedure values in real-time. The calculated KAP and CAK were compared to the values displayed by the fluoroscopy machine with excellent agreement. The DTS now is able to automatically calculate both KAP and CAK without the need for measurement by an add-on transmission ionization chamber.
机译:我们开发的皮肤剂量跟踪系统(DTS)在荧光检查过程中实时提供患者3D图形上累积皮肤剂量分布的彩色编码映射。现在已对DTS进行了修改,以使用从Toshiba Infinix系统上从数字总线实时获得的数据来计算荧光检查的比释动能面积乘积(KAP)和累积空气比释动能(CAK)。 KAP是整个光束区域的空气比释动能的整数倍,通常使用准直仪组件中包含的大面积透射电离室进行测量。在该软件中,对于给定的kVp和射束过滤乘以每个脉冲的mAs乘以射束的长度和宽度乘以一个场,从校准文件中自动得出每个X射线脉冲的KAP,作为空气比释动能/ mAs的乘积。不均匀校正因子。场的不均匀性主要是脚跟效应的结果,校正因子是根据使用射线变色胶片测得的光束轮廓确定的。将KAP除以介入参考点的射束面积即可得到面积平均的CAK。在每个脉冲之后,将每个X射线脉冲的KAP和CAK相加,以实时获取总过程值。将计算出的KAP和CAK与荧光透视仪显示的值进行比较,吻合得很好。 DTS现在能够自动计算KAP和CAK,而无需通过附加的传输电离室进行测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号