MSV signal processing system for neutron-gamma discrimination in a mixed field

机译:用于混合场中子伽玛分辨的MSV信号处理系统

获取原文

摘要

Based on the principles derived from Campbell’s theorem, this paper carries out an analysis of the possibilities of Campbell’s mean square value signal processing system. The mean square value mode is especially suitable for measurements performed in a mixed radiation field, because the quantities of electrical charge involved in the interactions of the two types of radiation are substantially different. The measuring detector element may be an adequate ionization chamber and/or semiconductor components for mixed n-γ fields. An examination of the discrimination of gamma in relation to the neutron component in the signal of the detector output was carried out, calculated according to the theoretical model of radiation interaction with the detector. The advantage of the mean square value method was confirmed and it was concluded that the order of n-γ discrimination in mean square value signal processing is greater than the one rendered by the classical measuring method.
机译:基于从坎贝尔定理得出的原理,本文对坎贝尔均方信号处理系统的可能性进行了分析。均方值模式特别适合在混合辐射场中执行的测量,因为参与两种类型的辐射相互作用的电荷量基本上不同。测量检测器元件可以是用于混合的n-γ场的足够的电离室和/或半导体部件。根据探测器与辐射相互作用的理论模型,计算了探测器相对于探测器输出信号中子成分的伽玛辨别力。证实了均方值方法的优点,并得出结论,均方值信号处理中n-γ判别的阶数大于经典测量方法所给出的阶数。