Scanning Electron Micrographic Investigation of Explanted (Used) Metallic Biomedical Implants for Corrosion and Other Mechanical Damage Responsible for Failure

机译:植入的(使用过的)金属生物医学植入物的扫描电子显微照相研究,其负责造成故障的腐蚀和其他机械损坏

获取原文
           

摘要

This paper deals with detailed corrosion analysis of explanted devices. The study of total 6 different types of orthopedic metallic implant was carried out after collecting the clinical report from the doctors, who performed these implantations. The clinical report covered the purpose of implantation, body part where implantation was done, and physiological reasons of removal of implant. The metallurgical investigation to study corrosion and any other mechanical damage to the implant surface during their service period was done using the Scanning Electron Micrography. SEM presented in this paper reveals the presence of in-vitro corrosion and mechanical damage as well, which are corroborating well with clinical reports.
机译:本文讨论了植入设备的详细腐蚀分析。在收集了进行这些植入手术的医生的临床报告后,对总共6种不同类型的整形外科金属植入物进行了研究。临床报告涵盖了植入的目的,进行植入的身体部位以及移除植入物的生理原因。使用扫描电子显微镜进行了冶金学研究,以研究腐蚀和植入物表面在使用期间的任何其他机械损伤。本文介绍的SEM揭示了体外腐蚀和机械损伤的存在,与临床报告很好地证实了这一点。