首页> 外文期刊>Korean journal of radiology : >Introducer Curving Technique for the Prevention of Tilting of Transfemoral Günther Tulip Inferior Vena Cava Filter
【24h】

Introducer Curving Technique for the Prevention of Tilting of Transfemoral Günther Tulip Inferior Vena Cava Filter

机译:防止弯曲股骨Günther郁金香下腔静脉滤器倾斜的引入弯曲技术

获取原文

摘要

Objective To determine whether the introducer curving technique is useful in decreasing the degree of tilting of transfemoral Tulip filters. Materials and Methods The study sample group consisted of 108 patients with deep vein thrombosis who were enrolled and planned to undergo thrombolysis, and who accepted transfemoral Tulip filter insertion procedure. The patients were randomly divided into Group C and Group T. The introducer curving technique was Adopted in Group T. The post-implantation filter tilting angle (ACF) was measured in an anteroposterior projection. The retrieval hook adhering to the vascular wall was measured via tangential cavogram during retrieval. Results The overall average ACF was 5.8 ± 4.14 degrees. In Group C, the average ACF was 7.1 ± 4.52 degrees. In Group T, the average ACF was 4.4 ± 3.20 degrees. The groups displayed a statistically significant difference (t = 3.573, p = 0.001) in ACF. Additionally, the difference of ACF between the left and right approaches turned out to be statistically significant (7.1 ± 4.59 vs. 5.1 ± 3.82, t = 2.301, p = 0.023). The proportion of severe tilt (ACF ≥ 10°) in Group T was significantly lower than that in Group C (9.3% vs. 24.1%, χ2 = 4.267, p = 0.039). Between the groups, the difference in the rate of the retrieval hook adhering to the vascular wall was also statistically significant (2.9% vs. 24.2%, χ2 = 5.030, p = 0.025). Conclusion The introducer curving technique appears to minimize the incidence and extent of transfemoral Tulip filter tilting.
机译:目的确定导引管弯曲技术是否可用于减少经股郁金香滤器的倾斜程度。材料和方法本研究样本组由108名深静脉血栓形成患者组成,他们均已入组并计划进行溶栓,并接受经股动脉Tulip滤器插入手术。将患者随机分为C组和T组。T组采用导引器弯曲技术。在前后投影中测量植入后滤器的倾斜角(ACF)。在切取过程中,通过切向腔图对附着在血管壁上的切取钩进行测量。结果总体平均ACF为5.8±4.14度。在C组中,平均ACF为7.1±4.52度。在T组中,平均ACF为4.4±3.20度。各组的ACF差异有统计学意义(t = 3.573,p = 0.001)。此外,左进路和右进路之间的ACF差异在统计上也很显着(7.1±4.59 vs. 5.1±3.82,t = 2.301,p = 0.023)。 T组严重倾斜(ACF≥10°)的比例显着低于C组(9.3%对24.1%,χ 2 = 4.267,p = 0.039)。在各组之间,取回钩附着在血管壁上的比率差异也具有统计学意义(2.9%比24.2%,χ 2 = 5.030,p = 0.025)。结论导管弯曲技术似乎使经股郁金香滤器倾斜的发生率和程度降到最低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号