Comparison of contamination between conventional three-way stopcock and needleless injection device: A randomized controlled trial

机译:传统三通旋塞阀和无针注射装置之间的污染比较:一项随机对照试验

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background:Intraluminal contamination of catheter hubs has been recognized as the most frequent cause of catheter-related blood stream infections. We have investigated the efficacy of a new hub device, Planecta SC? (PNSC), in preventing endoluminal catheter contamination, compared to a conventional three-way stopcock.Material/Methods: Adults patients requiring an intravascular catheter placement for at least 48 hours in intensive care units were randomly assigned to receive either the infusion device with the newly designed hub, PNSC (P group, n=89), or with a conventional three-way stopcock (C group, n=73). To evaluate intraluminal contamination, we examined the bacteria isolated in the inline bacterial filters which were attached to downstream of the injection ports. In addition to the clinical study, we conducted a bench study to investigate if use of protection caps or strict disinfection technique prevented intraluminal contamination with this new needleless connector.Results: The incidence of bacterial contamination was not significantly different between the groups (P group 9/89 (10.1%) vs. C group 6/73 (8.2%), P=0.79). There was no correlation between the numbers of injections, duration of the use of the device or the microbial contamination rate. In the bench study, protection caps and disinfection technique significantly decreased microbial transfer from the hub to the fluid space.Conclusions:We concluded that the use of the new hub device did not reduce endoluminal bacterial contamination rate in comparison with that of a three way stopcock. Intraluminal bacterial contamination may be reduced by either strict disinfection technique or when a protection cap is use.
机译:背景:导管座的腔内污染已被认为是与导管相关的血流感染的最常见原因。我们已经研究了新型集线器设备Planecta SC的功效。 (PNSC),与传统的三通旋塞阀相比,在预防腔内导管污染方面。材料/方法:将需要在重症监护病房中放置血管内导管至少48小时的成人患者随机分配到接受输液装置的同时新设计的球阀PNSC(P组,n = 89),或使用传统的三通旋塞阀(C组,n = 73)。为了评估腔内污染,我们检查了在线细菌过滤器中分离出的细菌,这些细菌附着在注入口的下游。除了临床研究外,我们还进行了一项台式研究,以研究使用这种新的无针连接器使用保护帽或严格的消毒技术是否可以防止腔内污染。结果:两组之间的细菌污染发生率无显着差异(P组9 / 89(10.1%)vs.C组6/73(8.2%),P = 0.79)。注射次数,使用设备的持续时间或微生物污染率之间没有相关性。在基准研究中,保护帽和消毒技术显着减少了微生物从集线器到液体空间的转移。结论:我们得出的结论是,与三通旋塞阀相比,使用新的集线器装置不会降低腔内细菌污染率。 。通过严格的消毒技术或使用保护盖可以减少腔内细菌污染。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号