首页> 外文期刊>Medical Devices: Evidence and Research >Thyroid surgery with a harmonic scalpel: an experimental study
【24h】

Thyroid surgery with a harmonic scalpel: an experimental study

机译:谐波手术刀进行甲状腺手术的实验研究

获取原文
       

摘要

Background: The goal of the study was to determinate the safety of the harmonic scalpel, widely used in thyroidectomy, near the recurrent laryngeal nerve (RLN). Methods: The study involved ten pigs of either sex. Twenty RLNs at risk were dissected using the new harmonic scalpel FOCUS. The distances between the nerve and the activated instrument were checked with a millimeter ruler. After dissection, the pigs were euthanized, and both RLNs were fixed in formol and examined by histology after staining with hematoxylin–eosin. Due to technical reasons, only 18 RLNs from the ten pigs could be examined. Results: In the experiment that investigated the extent of heat injury, ultrasonic dissection did not cause any immediate damage of the nerve even close to the RLN (1 mm away from the RLN). Conclusion: The use of harmonic scalpel FOCUS for thyroid surgery is safe for the surrounding structures (nerves). Careful tissue applications of the device near the RLN (1 mm) did not cause any lesion histologically.
机译:背景:该研究的目的是确定在喉返神经(RLN)附近广泛用于甲状腺切除术的谐波手术刀的安全性。方法:该研究涉及十只两性猪。使用新的谐波解剖刀FOCUS解剖了20个有风险的RLN。用毫米尺检查神经与被激活器械之间的距离。解剖后,对猪实施安乐死,将两个RLN固定在甲醛中,并用苏木精-伊红染色后进行组织学检查。由于技术原因,只能检查10头猪的18个RLN。结果:在研究热损伤程度的实验中,即使在靠近RLN(距RLN 1毫米)的地方,超声解剖也不会对神经造成任何直接损害。结论:谐波解剖刀FOCUS用于甲状腺手术对周围结构(神经)是安全的。小心地在RLN(1 mm)附近组织应用该装置在组织学上未引起任何病变。