A Biomechanical Comparison of Shape Design and Positioning of Transforaminal Lumbar Interbody Fusion Cages

机译:经椎间孔腰椎椎间融合器笼的形状设计和定位的生物力学比较

获取原文
       

摘要

Study Design Cadaveric biomechanical analysis. Objective The aim of this study was to compare three interbody cage shapes and their position within the interbody space with regards to construct stability for transforaminal lumbar interbody fusion. Methods Twenty L2–L3 and L4–L5 lumbar motion segments from fresh cadavers were potted in polymethyl methacrylate and subjected to testing with a materials testing machine before and after unilateral facetectomy, diskectomy, and interbody cage insertion. The three cage types were kidney-shaped, articulated, and straight bullet-shaped. Each cage type was placed in a common anatomic area within the interbody space before testing: kidney, center; kidney, anterior; articulated, center; articulated, anterior; bullet, center; bullet, lateral. Load-deformation curves were generated for axial compression, flexion, extension, right bending, left bending, right torsion, and left torsion. Finally, load to failure was tested. Results For all applied loads, there was a statistically significant decrease in the slope of the load-displacement curves for instrumented specimens compared with the intact state ( p p =?0.062). Among all instrumented groups, there was no statistically significant difference in stiffness for any of the loading conditions or load to failure. Conclusions Our results failed to show a clearly superior cage shape design or location within the interbody space for use in transforaminal lumbar interbody fusion.
机译:研究设计尸体生物力学分析。目的本研究的目的是比较三种椎间融合器的形状及其在椎间空间中的位置,以构建经椎间孔腰椎椎间融合的稳定性。方法将新鲜尸体的20个L2-L3和L4-L5腰椎运动段装入聚甲基丙烯酸甲酯中,并在单侧小平面切除术,椎间盘切除术和椎间融合器插入前后,用材料测试仪进行测试。三种笼子类型为肾形,铰接形和直子弹形。在测试之前,将每种笼子类型放置在椎体间空间的共同解剖区域中:肾脏,中心;肾脏,前部;中心铰接关节前子弹,中心;子弹,侧面。产生了载荷-变形曲线,用于轴向压缩,弯曲,延伸,右弯曲,左弯曲,右扭转和左扭转。最后,测试了失败的负载。结果对于所有施加的载荷,与完整状态相比,仪器化试样的载荷-位移曲线的斜率有统计上的显着下降(p p =?0.062)。在所有仪器组中,对于任何载荷条件或破坏载荷,刚度在统计学上均无统计学差异。结论我们的结果未能显示出用于椎间孔椎间融合术的明显优越的笼形设计或椎体间位置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号