Comparison of the VersaTrek and BACTEC MGIT 960 systems for the contamination rate, time of detection and recovery of mycobacteria from clinical specimens

机译:比较VersaTrek和BACTEC MGIT 960系统的污染率,从临床标本中检测和恢复分枝杆菌的时间

获取原文
       

摘要

The most commonly used conventional Lowenstein Jensen (L.J) culture method requires at least 6 to 10 weeks of incubation due to the slow growth rate of the?Mycobacterium tuberculosis?complex. Today, there are liquid automated systems to detect mycobacteria than faster conventional LJ culture method. In this study, we aimed to compare of the VersaTrek system and BACTEC MGIT 960 system for the contamination rate and recovery of mycobacteria from clinical specimens. The recovery rates of mycobacteria were detected as 4% (2/50) for both of the VersaTrek and MGIT 960 systems. No statistically significant difference was detected for the recovery rates of mycobacteria in both of the systems (p>0.05). Two strains were detected on 12 and 15 day in MGIT 960 system. Two strains were detected on 15 and 20 day in VersaTrek system. No statistically significant difference was detected for the time of detection of mycobacteria in both of the systems (p>0.05). The contamination rates were found as 10% (5/50) and 6% (3/50) in MGIT 960 and VersaTrek, systems respectively (p>0.05). No statistically significant difference was detected for the contamination rates for both of the systems. Antituberculosis resistance tests were performed using with antituberculosis agents in MGIT 960 and VersaTrek systems. The two strains were found sensitive against to streptomycin, isoniazid, rifampicin and ethambutol in MGIT 960 system. The two strains were found sensitive against to isoniazid, rifampicin and ethambutol in VersaTrek system. Streptomycin was not used in VersaTrek system because there is no standardization for streptomycin. We also examined the time of detection of 5 confirmedM.?tuberculosis?complex members in MGIT 960 and VersaTrek systems. When the time of detection in MGIT 960 and VersaTrek systems were compared, no statistically significant difference was found (p>0.05).?In conclusion, no substantial difference has been observed between MGIT 960 and VersaTrek systems for?M. tuberculosis?isolation and contamination rate. We believe that VersaTrek media, which is going to be put into use in Turkey, is easy for application and follow up, gives rapid and sensitive results and that it can be used in routine alike MGIT media.
机译:由于“结核分枝杆菌”复合体的缓慢生长,最常用的常规Lowenstein Jensen(L.J)培养方法需要至少培养6至10周。如今,有比自动LJ培养方法更快的液体自动化系统可以检测分枝杆菌。在这项研究中,我们旨在比较VersaTrek系统和BACTEC MGIT 960系统在临床标本中的感染率和分枝杆菌的回收率。对于VersaTrek和MGIT 960系统,分枝杆菌的回收率均为4%(2/50)。在两个系统中,分枝杆菌的回收率均未发现统计学上的显着差异(p> 0.05)。在MGIT 960系统中,在第12天和第15天检测到两个菌株。在15天和20天在VersaTrek系统中检测到两个菌株。在两个系统中检测分枝杆菌的时间均未检测到统计学上的显着差异(p> 0.05)。在MGIT 960和VersaTrek系统中,污染率分别为10%(5/50)和6%(3/50)(p> 0.05)。两个系统的污染率均未检测到统计学上的显着差异。在MGIT 960和VersaTrek系统中使用抗结核药进行抗结核药耐药性测试。在MGIT 960系统中发现这两种菌株对链霉素,异烟肼,利福平和乙胺丁醇敏感。在VersaTrek系统中,发现这两种菌株对异烟肼,利福平和乙胺丁醇敏感。 VersaTrek系统中未使用链霉素,因为链霉素尚无标准化方法。我们还检查了在MGIT 960和VersaTrek系统中检测到的5个已确诊的结核分枝杆菌复杂成员的时间。当比较MGIT 960和VersaTrek系统的检测时间时,没有发现统计学上的显着差异(p> 0.05)。总之,在MGIT 960和VersaTrek系统之间,没有发现实质性差异。结核病的隔离和污染率。我们认为,将在土耳其使用的VersaTrek媒体易于应用和跟进,可提供快速而敏感的结果,并且可以像MGIT媒体一样用于常规。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号