Comparison of PET/CT and PET/MR imaging and dosimetry of yttrium-90 ( 90 Y) in patients with unresectable hepatic tumors who have received intra-arterial radioembolization therapy with 90 Y microspheres

机译:接受90 Y微球动脉内放射栓塞治疗的无法切除的肝肿瘤患者的PET / CT和PET / MR成像以及yttrium-90(90 Y)剂量的比较

获取原文
           

摘要

Abstract BackgroundThe aim of our study was to compare 90Y dosimetry obtained from PET/MRI versus PET/CT post-therapy imaging among patients with primary or metastatic hepatic tumors.First, a water-filled Jaszczak phantom containing fillable sphere with 90Y-chloride was acquired on both the PET/CT and PET/MRI systems, in order to check the cross-calibration of the modalities. Following selective internal radiation therapy (SIRT) with 90Y microspheres, 32 patients were imaged on a PET/CT system, immediately followed by a PET/MRI study. Reconstructed images were transferred to a common platform and used to calculate 90Y dosimetry. A Passing-Bablok regression scatter diagram and the Bland and Altman method were used to analyze the difference between the dosimetry values.ResultsThe phantom study showed that both modalities were calibrated with less than 1% error. The mean liver doses for the 32 subjects calculated from PET/CT and PET/MRI were 51.6?±?24.7?Gy and 46.5?±?22.7?Gy, respectively, with a mean difference of 5.1?±?5.0?Gy. The repeatability coefficient was 9.0 (18.5% of the mean). The Spearman rank correlation coefficient was very high, ρ =?0.97. Although the maximum dose to the liver can be significantly different (up to 40%), mean liver doses from each modalities were relatively close, with a difference of 18.5% or less.ConclusionsThe two main contributors to the difference in 90Y dosimetry calculations using PET/CT versus PET/MRI can be attributed to the differences in regions of interest (ROIs) and differences attributed to attenuation correction. Due to the superior soft-tissue contrast of MRI, liver contours are usually better seen than in CT images. However, PET/CT provides better quantification of PET images, due to better attenuation correction. In spite of these differences, our results demonstrate that the dosimetry values obtained from PET/MRI and PET/CT in post-therapy 90Y studies were similar.
机译:摘要背景本研究的目的是比较原发性或转移性肝肿瘤患者从PET / MRI和PET / CT治疗后影像学获得的90Y剂量。在PET / CT和PET / MRI系统上,以检查模态的交叉校准。在对90Y微球进行选择性内部放射治疗(SIRT)之后,将32位患者在PET / CT系统上成像,然后立即进行PET / MRI研究。重建的图像被转移到一个通用平台,并用于计算90Y剂量。使用Passing-Bablok回归散点图以及Bland和Altman方法分析剂量学值之间的差异。结果幻像研究表明,两种模态均已校准,误差小于1%。根据PET / CT和PET / MRI计算的32位受试者的平均肝剂量分别为51.6±±24.7μGy和46.5±±22.7μGy,平均差异为5.1±±5.0μGy。重复性系数为9.0(平均值的18.5%)。 Spearman等级相关系数非常高,ρ=?0.97。尽管对肝脏的最大剂量可能有显着差异(最高40%),但每种方式的平均肝脏剂量相对接近,相差18.5%或更少。 / CT与PET / MRI的关系可以归因于关注区域(ROI)的差异,也归因于衰减校正。由于MRI具有出色的软组织对比度,通常比CT图像更能清晰显示肝脏轮廓。然而,由于更好的衰减校正,PET / CT提供了更好的PET图像量化。尽管存在这些差异,我们的结果表明,在90Y治疗后的研究中,从PET / MRI和PET / CT获得的剂量学值相似。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号