首页> 外文期刊>International Journal of Pharmaceutical Sciences and Research >TRANSMISSION & LEAKAGE MEASUREMENT OF NOVEL TELECOBALT MACHINE BHABHATRON-II
【24h】

TRANSMISSION & LEAKAGE MEASUREMENT OF NOVEL TELECOBALT MACHINE BHABHATRON-II

机译:新型百事达-II型电传机的传输和泄漏测量

获取原文

摘要

The Telecobalt machines are extensively used in developing countries due to its lesser cost, lesser power consumption and easy maintenance. The high activity source can be extensively used for at least 6-8 years due to its long half-life. Installing a Telecobalt unit comprises of fixing the machine in the treatment room, ensuring its mechanical motions, source loading and all other related Quality Assurance Checks. The Bhabhatron-II, an indigenous Telecobalt machine is installed at Acharya Tulsi Regional Cancer Centre, produced in India by Department of Atomic Energy, Mumbai. This paper describes about the installation & Commissioning of Bhabhatron-II and their related Quality Checks. The necessary Radiation Quality Assurance checks such as Combined Jaw Transmission, Individual Jaw Transmission, and ON Position Leakage in patient plane, Leakage in other than patient plane are measured. The collimator transmission of combined jaw and, Individual jaws (X & Y) are measured and they are 0.024%, 0.347% & 0.635% of the useful beam respectively. The ON position leakage in patient plane is measured by forming an imaginary circle of 2m radius @ SAD; the leakage level is 0.005% which is 5.42% of the recommended limit by I.E.C. The ON Position leakage level in other than patient plane is measured by a special apparatus which partially simulates a sphere of 1m radius. The leakage level is 0.347% of the Dose Rate measured at 1m. The Dosimetric instruments which are available in routine radiotherapy department are used. The TLD Disks are used to measure the ON Position Leakage in other than Patient plane. The measured readings are well under the Tolerance recommended by A.E.R.B India. The machine has been approved by the regulatory board and dedicated for Treatment purpose.
机译:Telecobalt机器因其成本较低,功耗较小且易于维护而在发展中国家广泛使用。高活性源由于其较长的半衰期,可以广泛使用至少6-8年。安装Telecobalt装置包括将机器固定在治疗室中,确保其机械运动,源装载和所有其他相关的质量保证检查。 Bhabhatron-II是一台本地钴钴机,安装在孟买原子能部门在印度生产的Acharya Tulsi地区癌症中心。本文介绍了Bhabhatron-II的安装和调试及其相关的质量检查。测量必要的辐射质量保证检查,例如组合下颌透射,单独下颌透射以及患者平面中ON位置泄漏,患者平面以外的泄漏。测量了组合钳口和单个钳口(X和Y)的准直仪透射率,它们分别为有用光束的0.024%,0.347%和0.635%。通过在SAD上形成2m半径的假想圆来测量患者平面中ON位置的泄漏;泄漏水平为0.005%,是I.E.C建议的极限的5.42%。除患者平面外,ON Position泄漏水平是通过特殊设备测量的,该设备部分模拟了1m半径的球体。泄漏量为剂量在1m时的0.347%。使用常规放射治疗部门中可用的剂量仪。 TLD磁盘用于测量患者平面以外的ON位置泄漏。测得的读数完全在印度A.E.R.B建议的公差范围内。该机器已通过监管委员会的批准,专门用于治疗目的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号