首页> 外文期刊>International Journal of Medical Physics, Clinical Engineering and Radiation Oncology >Direct Measurement of Medical Linear Accelerator Electron Beam Width at Scattering Foil Position
【24h】

Direct Measurement of Medical Linear Accelerator Electron Beam Width at Scattering Foil Position

机译:散射箔位置上医用直线加速器电子束宽度的直接测量

获取原文
           

摘要

Purpose: The aim of this study was to develop a method for the direct measurement of electron beam width and distribution at the scattering foil on the carrousel in a medical linear accelerator gantry head, which differs from an existing indirect method for measuring the focal spot size using a camera or metallic slit located outside the gantry head. Methods: The electron beam emitted by the linear accelerator was used to irradiate radiochromic film mounted on the scattering foil on the carrousel, which was not used for clinical treatment. The electron beam width at the scattering foil position was then evaluated using the full width at half maximum of the Gaussian distribution approximated from each one dimensional distribution of the irradiated radiochromic film. Results: The electron beam width at the scattering foil position was found to be 3.1 to 6.4 mm in the crossline direction and 2.8 to 5.5 mm in the inline direction with electron energy of 4 to 16 MeV. The two-dimensional distribution of the electron beam was therefore elliptical or distorted in shape, not circular. Conclusions: Direct measurement of the electron beam width at the scattering foil in the carrousel of a medical linear accelerator is possible, though the use of lower sensitivity film in addition to indirect methods is expected to bring about better results. However, as this method does not allow for direct measurement of the incident angle of the accelerated electron beam, further improvements and refinements are still needed.
机译:目的:本研究的目的是开发一种直接测量电子直线加速器龙门架中转盘上散射片上电子束宽度和分布的方法,该方法不同于现有的间接测量焦点尺寸的方法使用位于机架头外部的摄像头或金属狭缝。方法:线性加速器发射的电子束用于辐照安装在转盘上散射箔上的放射致变色膜,该膜不用于临床。然后使用高斯分布的半峰全宽估算散射箔位置的电子束宽度,高斯分布的半峰全宽近似于所辐射的辐射致变色膜的每个一维分布。结果:在散射箔位置处的电子束宽度在交叉线方向上为3.1至6.4mm,在行内方向上为2.8至5.5mm,电子能量为4至16MeV。因此,电子束的二维分布为椭圆形或变形的形状,而不是圆形的。结论:可以直接测量医用线性加速器转盘中散射箔处电子束的宽度,尽管除了间接方法外,使用较低灵敏度的薄膜有望带来更好的结果。但是,由于该方法不能直接测量加速电子束的入射角,因此仍需要进一步的改进和改进。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号