Study of Image Contrast Recovery Coefficient (CRC) of a Large Ring PET Scanner

机译:大型环形PET扫描仪图像对比度恢复系数(CRC)的研究

获取原文

摘要

Image contrast recovery coefficient (CRC) of a large ring PET scanner “macroPET” was studied with septa and without septa configuration by ac-quiring data from a laboratory made 35.7 cm square phantom filled with 18-F. Images were reconstructed with simple 2D filtered back projection using Hann, Hamming and Parsen filters with different cut-off frequencies aiming to investigate the influence of filter and cut-off frequency on image contrast with septa and without septa mode. Results indicate that the CRC, for both hot and cold lesions, is excellent for diameters ≥ 3 cm using cut-off frequencies > 0.4. For a 2 cm hot lesion CRC is around 0.8 to 0.9. CRC for 1 cm hot and cold lesions is ~0.3, as expected. There is surprisingly little difference between results with and without septa. For hot lesions, septa appear to improve CRC slightly, but for cold lesions CRC is slightly poorer using septa.
机译:通过获取实验室制造的填充了18-F的35.7平方厘米幻像的数据,研究了具有隔片和无隔片配置的大型环形PET扫描仪“ macroPET”的图像对比度恢复系数(CRC)。使用具有不同截止频率的Hann,Hamming和Parsen滤波器,通过简单的2D滤波反投影来重建图像,旨在研究具有隔片和无隔片模式的滤波器和截止频率对图像对比度的影响。结果表明,对于热损伤和冷损伤,CRC均适用于直径≥3 cm且截止频率> 0.4的情况。对于2厘米的热病变,CRC约为0.8至0.9。如预期的那样,1 cm冷热损伤的CRC约为0.3。有隔垫和无隔垫的结果之间几乎没有令人惊讶的差异。对于热损伤,隔垫似乎可以稍微改善CRC,但是对于冷损伤,使用隔垫时CRC稍微较差。